background image

A.  Introduction 

3                                     Questions about eRapid?   Call 1-866-335-6943 

Table of Contents 

  Quick Instructions 

   i 

A Introduction 

  3 

B Safety Precautions 

  4 

C Getting Started 

  5 

D Using Batteries or the AC Power Supply   6 
E  How to Clean and Maintain Your eRapid   7 
F  How to Assemble Your Handset 

13 

G How to Take Your Treatment 

15 

H Frequently Asked Questions 

18 

I  Troubleshooting 

19 

J  Technical Data and Symbols 

21 

K Warranty and Disposal  

22 

 

Pour le français voir: 

www.pari.com/us-en/products

 

Para español ver: 

www.pari.com/us-en/products

 

 

 

 

A.  Introduction

 

The eRapid® Nebulizer System (eRapid) is a lightweight, battery 

operated electronic nebulizer which will convert your liquid medicine 

into a fine mist. Your medicine is placed in a reservoir, which feeds to 

a stainless steel membrane that has been drilled with very small 

holes.  When this membrane is vibrated, liquid is forced through the 

holes to form a fine mist.

 

 

Before using your eRapid Nebulizer System, read and understand all 

Instructions For Use.  Save these Instructions For Use for future 

reference.  If for any reason you 

Do Not

 understand any part of these 

Instructions For Use, please call the toll-free number below before 

proceeding with your treatment.   
Take special note of all safety precautions marked 

Danger, Warning,  

Caution or Important.

 

Safety Precaution Definitions: 
Danger 

indicates hazards, which, without precautionary measures, 

will result in death or serious injury. 

Warning 

indicates hazards, which, without precautionary measures, 

can result in serious injury or even death. 

Caution 

indicates hazards, which, without precautionary measures, 

can result in moderate or minor injury or impair treatment, or damage 

to equipment. 

Important

 indicates information considered important but not hazard 

related. 

Indications For Use 

The eRapid Nebulizer System is to be used with patients for whom 

doctors have prescribed medicine for nebulization. It is intended for 

adult and pediatric patients, and may be used in hospitals, hospital-

type facilities, nursing homes, sub-acute institutions and home 

environments. 

Содержание eRapid

Страница 1: ...General Purpose Electronic Nebulizer...

Страница 2: ...ng Handset components in warm soapy water for 5 minutes 3 Rinse all parts 4 Disinfect Handset components by boiling in distilled water for 5 minutes 5 Air dry Handset components until dry 3 Assemble a...

Страница 3: ...lean by soaking Handset components in warm soapy water for 5 minutes 3 Rinse all components to remove residual soap and inspect for dirt Soak for an additional 5 minutes if necessary 4 Disinfect by bo...

Страница 4: ...d Nebulizer System read and understand all Instructions For Use Save these Instructions For Use for future reference If for any reason you Do Not understand any part of these Instructions For Use plea...

Страница 5: ...ow when the Controller is damaged or not properly assembled 12 To reduce the risk of infection illness and or injury from contamination and to maintain the longevity of your Handset clean and dry all...

Страница 6: ...be cleaned and or disinfected before you use it for the first time Remove one new Handset and one new Aerosol Head from their packaging and proceed to step E2c to begin the cleaning procedure You can...

Страница 7: ...output Operating life is about 2 hours when using high quality disposable batteries To reduce the risk of battery leakage it is recommended to remove the batteries if you plan to store the Controller...

Страница 8: ...ishes Use only distilled water to disinfect by boiling Follow manufacturer directions when using a chemical disinfectant such as Control III Handset Maintenance Summary Details Can Be Found on the Fol...

Страница 9: ...rformance It is important to rinse and clean your Handset right away after every use Step E1 Disconnect the Handset from the Connection Cord Figure E1 Step E2 Disassemble the eRapid Handset a Remove t...

Страница 10: ...Figure E4 Never use a brush or any other object to clean the Aerosol Head Thorough rinsing of both sides of the Aerosol Head helps with proper operation and prevents clogging Step E5 Soak all parts of...

Страница 11: ...lled water for a full 5 minutes Figure E9b Step E10 Dry on a lint free cloth or paper towel Figure E10 After the parts are completely dry place them in a bag or cover them with a lint free cloth or pa...

Страница 12: ...ce Step E13 Connect the Easycare to the Controller using the Connection Cord by attaching the horseshoe shaped end of the Connection Cord upside down to the hinge of the Easycare part and swivel it do...

Страница 13: ...y pressing and holding the On Off button for 10 seconds Step E17 Reverse the assembly steps and discard any solution from the Easycare Step E18 Clean the Easycare by soaking in warm soapy water Step E...

Страница 14: ...re clean and not visibly damaged Do Not use dirty or damaged parts Step F3 Open the Handset The Medication Reservoir and the Aerosol Chamber are attached using a hinge First gently pull up on the tab...

Страница 15: ...it It is important that you Do Not touch the metal part of the Aerosol Head Figure F5c Step F6 Close the Handset by pushing the Aerosol Chamber together with the Medication Reservoir until you hear o...

Страница 16: ...id Do not exceed the maximum fill volume of 6 mL Step G4 Close the Medication Cap by lining up the Tabs on the Medication Cap with the Tab Slots on the Medication Reservoir and turning the Medication...

Страница 17: ...Flap with your mouth Step G8 Turn on the Controller Press and hold the On Off button for 1 2 seconds Figure G8a The following will happen You will hear 1 beep and the status light will turn green The...

Страница 18: ...opening Figure G11 To mimic other general purpose nebulizers the eRapid is designed to keep 1 mL 1 5 teaspoon of medicine in the Medication Reservoir Do not tilt or try to use the residual medicine If...

Страница 19: ...he Batteries when storing the Controller for more than 30 days The Handset should last for 2 6 months but it depends on the number of different medicines used with the Handset the number of times the...

Страница 20: ...s blinking No medicine Add medicine if you have not done so Not level or medicine is not touching the Aerosol Head Gently tap the Handset hold the Handset level and press the On Off button 7 Blinking...

Страница 21: ...ndition persists call the number below 16 Longer than normal nebulization time Clean the Aerosol Head and then use Easycare If condition persists call the number below 17 There is more than 1 mL resid...

Страница 22: ...01 1 Type of electric shock protection AC Power Supply Protection Class II Degree of protection from electric shock of part used nebulizer Type BF Degree of protection against water ingress per IEC 60...

Страница 23: ...ent should fail during the warranty period PARI at its option will repair or replace the equipment or parts at issue provided the claim is bona fide It shall be the responsibility of the purchaser to...

Страница 24: ...n Virginia 23112 USA 2019 PARI Pharma GmbH All rights reserved The PARI Pharma logo is a registered trademark of PARI GmbH Lamira eBase Altera Zirela eFlow and the eFlow TECHNOLOGY logo are registered...

Отзывы: