Customer Guide
HPAC Cell Cleaning
©2015 Parata Systems, LLC
page 35 of 40
Parata Technical Assistance Center
120-0465 Rev 03
1-888-PARATA1 (727-2821)
1.
Disassemble the chute assembly for cleaning:
1.1
Lift the arm on one side of the door away from the assembly. The two pins on
the arm will come out of the hole in the tube and the slot on the dispense
chute.you).
Step 1.1
1.2
With the pins out of the holes, release the arm and let it rest against the
assembly.
1.3
Lift the arm on the other side of the door away from the assembly to disconnect
the door, then remove the door.
1.4
Slide the tube out of the dispense chute.
2.
Clean and dry all three parts of the chute assembly using a pharmacy-approved method.
3.
Visually inspect each part to verify that all residue is removed and the part is dry.
•
If you see any moisture inside the chute assembly
, allow it to continue air-drying.
Do not
proceed until the chute assembly looks completely dry.
•
If you see any pill residue inside the chute assembly,
repeat the cleaning procedure
(Step 2). If you still see pill residue after the second cleaning (or if, for any reason, you