©2015 Parata Systems, LLC. All rights reserved. No part of this documentation or the associated 
product may be reproduced, processed, duplicated or published in any form by photocopying, 
microfilming, reprinting or other process, in particular electronic means, without the written 
agreement of Parata Systems, LLC.

Trademarks

AccuCount, AccuMed, APM, Automated Will Call, AutoScript III, Baker Cassette, Baker Cell, Baker 
Universal, Baker Universal 2010, FlexCheck, onePac, Parata, Parata and design, Parata 
Consulting, Parata Front Runners Club, Parata Express, Parata , Parata Max, Parata 
PASS 500, Parata PASS 208, Parata PASS Pack, Parata Check PASS, Parata First PASS, Parata 
PASS Ware, Parata PASS Safe Loader, Max, Max+, Max Flex, Mini, Circle of Pharmacy Care, Circle 
of Pharmacy Care Fill Serve Adhere and design, Next-Generation Pharmacist, Next Generation 
Pharmacist logo, Peak Pharmacy Performance, P3 Peak Pharmacy Performance and design, 
Parata Performance Services, Parata Uber Module, and Parata Puts You Out In Front are 
trademarks or registered trademarks of Parata Systems, LLC. Pharmacy 2000 is a trademark of 
McKesson Automation Systems, Inc. PACMED and PACVISION are trademarks of McKesson 
Automation Inc. in the US. PACMED and PACVISION are trademarks or registered trademarks of 
McKesson Automation Canada Corporation in Canada. GREENGUARD Indoor Air Quality 
Certified, UL and the UL logo are trademarks or registered trademarks of UL, LLC. Axeda and 
ServiceLink are trademarks or registered trademarks of Axeda Corporation. Microsoft and 
Windows are registered trademarks or trademarks of Microsoft Corporation in the United States 
and/or other countries. Other brands and their products are trademarks or registered 
trademarks of their respective holders and should be noted as such.

Patents

Refer to www.parata.com/patents-and-trademarks/ for patents owned by Parata Systems, LLC.

United States Government Rights

Use, duplication or disclosure by the U.S. government is subject to the restrictions set forth in 
the Parata Systems, LLC Agreement and as provided in DFARS 227.7202-1(a), 227.7202-3(a) 
and 227.7202-4 (1998 ed., revised June 21, 2005), DFARS 252.227-7013(b)(3) and (c) (NOV 
1995, revised January 23, 2006), FAR 12.212(a) (1995), FAR 52.227-19, or FAR 52.227-14 (ALT 
III), as applicable. Unpublished rights are reserved under the copyright laws of the United States.

Disclaimer

The information contained herein is considered accurate at the time of publication. Parata 
Systems, LLC reserves the right to make changes to the documentation or associated product at 
any time without notice.

Содержание Mini

Страница 1: ...r Guide 120 0465 Rev 03 Parata Systems LLC 2600 Meridian Parkway Durham NC 27713 www parata com The information in this document is proprietary and confidential to Parata Systems LLC Content may be ch...

Страница 2: ...son Automation Inc in the US PACMED and PACVISION are trademarks or registered trademarks of McKesson Automation Canada Corporation in Canada GREENGUARD Indoor Air Quality Certified UL and the UL logo...

Страница 3: ...ng with an established safety program pertaining to handling drugs and dispensing them with pharmacy automation Dispensing hazardous drugs with a Parata dispensing unit is not recommended For addition...

Страница 4: ...eting the cell drying process Reassembling the cell Replenishing the cell 1 Prepare the cell for cleaning Note Leave the unit on and the Mini applications running 1 1 Obtain a pair of gloves from the...

Страница 5: ...Close the cell door 1 7 Remove the pleated breather Step 1 7 Tab location on the breather 1 7 a Push the tab toward the nozzle end of the cell and lift up to remove the pleated breather from the cell...

Страница 6: ...ther is included in the kit 1 8 Remove the foam breather 1 8 a If there is tape securing the nozzle cover to the cell remove the tape and dispose of it according to pharmacy policy Note There may be t...

Страница 7: ...g the cell body with one hand push the index finger of your other hand through one of the center openings in the grid so that your finger depresses the foam breather and brace your thumb against the s...

Страница 8: ...nozzle cover and dispose of it according to pharmacy policy Note You will replace the foam breather you removed here with a new one later in the procedure A new foam breather is included in the cleani...

Страница 9: ...Push the cover firmly against the nozzle so that both sides of the cover snap into place over the tabs on the cell nozzle Step 1 9 c 1 10 If it is installed remove the foam plug next to the black hei...

Страница 10: ...nterclockwise to adjust the setting to E 0 1 10 d Dispose of the foam plug according to pharmacy policy Note You will replace the foam plug you removed here with a new one later in the procedure A new...

Страница 11: ...arata Technical Assistance Center 120 0465 Rev 03 1 888 PARATA1 727 2821 1 11 c Dispose of the baffle covers according to pharmacy policy Note You will replace the baffle covers you removed here with...

Страница 12: ...d toward the center of the nozzle to loosen the adhesive and lift the guard up and away from the cell Continue this process until you can grasp the guard and pull it away from the cell Step 1 12 a 1 1...

Страница 13: ...is easier to remove the adhesive when the cell is dry than when the cell is wet 1 14 a Use the sticky side of transparent tape to pull up any excess adhesive Step 1 14 a 1 14 b If necessary use a plas...

Страница 14: ...make sure that it is secure If it moves easily or comes off discard the stem cap and obtain another one from the kit 1 16 Verify that the cell is ready for cleaning Pills removed Pleated breather remo...

Страница 15: ...ld it and place it on the counter 2 2 Insert the cell into a zip closure bag from the cleaning kit 2 3 Pour one quart of Cleaning Solution red label into the bag Note One quart is the appropriate amou...

Страница 16: ...5 Rev 03 1 888 PARATA1 727 2821 2 5 Roll down the top of the bag This will make it easier to hold Step 2 5 2 6 Shake the cell vigorously for 1 minute Beginning on one side shake the cell front to back...

Страница 17: ...ce Center 120 0465 Rev 03 1 888 PARATA1 727 2821 2 7 Drain the excess liquid from the cell into the bag 2 7 a Hold the bag so that you can lift the cell by the door end and raise it out of the liquid...

Страница 18: ...2 9 Prepare the bag for disposal 2 9 a Pour half of a bottle of Sorbent Powder 20 grams into the liquid in the bag Within 1 to 2 minutes the liquid will congeal and become a white substance resemblin...

Страница 19: ...ell 3 1 Obtain a new drying cloth from the cleaning kit unfold it and place it in a location where the drying cell will not be disturbed for 24 hours 3 2 Insert the cell into a new zip closure bag fro...

Страница 20: ...so that you can lift the cell by the door end and raise it out of the liquid to drain 3 8 b Gently shake the cell to drain as much of the excess liquid as possible 3 9 Remove the cell from the bag and...

Страница 21: ...stem cap and dispose of it according to pharmacy policy Note The cap must be removed to allow air to flow through the cell Step 3 12 4 Air dry the cell for 24 hours 4 1 Position the cell so that the...

Страница 22: ...ry for 24 hours Step 4 3 Standard cell left and super cell right positioned door end down on the drying cloth 5 If applicable clean the chute assembly 5 1 Using the T handle driver from your starter k...

Страница 23: ...solenoid will retract Step 5 3 Chute Map with the icon unselected left and selected right 5 4 Pull the chute assembly out of the unit 5 5 Clean the chute assembly Note Refer to Appendix Cleaning the c...

Страница 24: ...Max and Mini Under certain conditions however it is possible that you still may see pill residue inside the cell even after the second cleaning If any residue remains in the cell or if for any reason...

Страница 25: ...kit and fold it to the appropriate size for this cell For a standard cell Fold the cloth into a 6 inch x 1 1 2 inch rectangle For a super cell Fold the cloth into a 12 inch x 1 1 2 inch rectangle Ste...

Страница 26: ...into the cell and position it so that it is under the baffle A and as far up toward the nozzle throat B as possible Step 6 6 Position of the drying cloth in a standard cell top and in a super cell bot...

Страница 27: ...remaining moisture 6 9 Visually inspect the cell If you see any moisture inside the cell allow it to continue air drying Do not proceed until the cell looks completely dry If the cell looks completely...

Страница 28: ...at the opening on the top of the cell then press down on the tab end until the breather snaps into place Step 7 2 b Inserting a new pleated breather 7 3 Install a new foam breather 7 3 a Obtain a new...

Страница 29: ...03 1 888 PARATA1 727 2821 7 3 d Hold the replacement foam breather in your other hand with the U shaped part of the foam breather pointed up and the wings pointed down Step 7 3 d 7 3 e Press the U sha...

Страница 30: ...the rounded end with the top of the cover facing up 7 3 h Slide the nozzle cover onto the nozzle so that the cutout on each side of the cover aligns with the corresponding tab on each side of the noz...

Страница 31: ...p between the nozzle and the body of the cell Then press the remaining tape down onto the side of the nozzle cover Step 7 4 b 7 4 c Use a flat object such as a counting spatula to press the tape again...

Страница 32: ...te the cell even if you are placing the same drug in the cell The settings will have changed during the cleaning process Follow the replenishment procedure in the Mini Operator s Guide to load pills i...

Страница 33: ...baffle covers on the cell one on each side of the baffle 8 1 a Position the baffle cover with the tab toward the top of the cell and place it over the baffle 8 1 b Apply pressure to the outer edge of...

Страница 34: ...ng the chute Parts The chute assembly consists of three parts the dispense chute A the door B and the tube C There are four 4 pins on the inside of the door arms that connect the assembly parts The tw...

Страница 35: ...ly 1 3 Lift the arm on the other side of the door away from the assembly to disconnect the door then remove the door 1 4 Slide the tube out of the dispense chute 2 Clean and dry all three parts of the...

Страница 36: ...ss of the chute assembly discard the chute assembly and order a new chute assembly If the chute assembly looks completely dry continue to Step 4 4 Reassemble the chute 4 1 Hold the dispense chute in y...

Страница 37: ...he channels on the dispense chute A Slide the tube forward until the hole in the rail lines up with the lower end of the angled slot C in the dispense chute Step 4 2 4 3 Hold the door in your dominant...

Страница 38: ...he arm into the hole in the tube rail The pin in the center of the arm will fit into the slot on the dispense chute Step 4 5 4 6 On the opposite side of the assembly lift the door arm and insert the p...

Страница 39: ...Customer Guide HPAC Cell Cleaning 2015 Parata Systems LLC page 39 of 40 Parata Technical Assistance Center 120 0465 Rev 03 1 888 PARATA1 727 2821 This page intentionally left blank...

Страница 40: ...Customer Guide HPAC Cell Cleaning 2015 Parata Systems LLC page 40 of 40 Parata Technical Assistance Center 120 0465 Rev 03 1 888 PARATA1 727 2821...