PGC, Inc.
Appendix B; SmartPad version 0.503
Steady State & Programmable
November 2005
B-27
ADDITIONAL PROGRAMMING INFORMATION
If you enter 00 00 00 for time in any segment, the SmartPad™ will stop at, but not
perform that segment. The time display will stop counting down, this is the indication that the
SmartPad™ is in the "halt" mode. Press "RUN" to exit the "halt" mode and continue with the
program.
If a segment must be longer than 17 hours, 59 minutes and 59 seconds for a ramp or
dwell, the time must be spread out over more than one segment.
If power to the SmartPad™ is removed for less than ~20 minutes, a program cycle will
resume from the point it was at when the unit was turned off or the power failed. If power is
removed for more than ~20 minutes the SmartPad™ will restart as defined on the "PROGRAM
INITIALIZATION" screen.