9
Getting Started with VISULEX Maki
VISULEX Maki
Getting Started with VISULEX Maki
Installation and Operation
Please position the base plate on a plain surface at your
working place. Release the thumbscrew. Place the
camera arm into the base plate so that the camera head
will be located above the base. Please fix the camera arm
by means of the thumbscrew.
Connectors at the Camera Arm
Please position the control panel in a way that you can
reach it without any difficulties. Connect the PS/2 cable
of the control panel to the PS/2 port at the camera arm.
Note: Do connect only the control panel which is
included in the delivery scope to the camera arm.
Other connected devices may lead to irreparable
damages.
Connect monitor and camera arm by means of a DVI-
cable.
Finally, plug the power adapter into a power socket and
place the connector into the port at the camera arm.
Now, your VISULEX Maki is ready to be used and can be
switched on.