TROUBLE SHOOTING
If the unit malfunctions, check out the following before calling for service.
Malfunction Check/Remedy
The chamber is not cooled
at all
The circuit breaker of power source is active
The voltage is too low. (In this case, call an electrician.)
The power switch is not ON.
The large amount of articles (load) is stored in the chamber at
one time.
The freezer is in defrost condition.
The cooling is poor
The ambient temperature is too high.
The outer door is not closed firmly.
The large amount of frost is built on the chamber wall.
The air intake vent is blocked.
The set temperature is not inputted properly.
The freezer is in the direct sunlight.
There is any heating source near the freezer.
A rubber cap and insulation for the access port are not set
correctly.
You put too many unfrozen articles into the freezer compartment.
Noise
The freezer is not installed on the sturdy floor.
The freezer is not leveled with the leveling feet.
There is anything touching the frame.
The freezer is in the status immediately after start up.
The unit sometimes causes a noise when the chamber temperature
is high due to the large load. The noise gets less and less
accompanying with the cooling of the chamber.
Note:
If the malfunction is not eliminated after checking the above items, or the malfunction is not shown in the
above table, contact our sales representative or agent.