background image

 

2

 

 INTRODUCTION

 

 
   Read this operating instructions carefully before using the appliance and follow the instructions for safety 
operation. 
 
   Our company never guarantee any safety if the appliance is used for any objects other than intended use 
or used by any procedures other than those mentioned in this operating instructions. 
 
   Keep this operating instructions in an adequate place to refer to it as necessary. 
 
   The contents of the operating instructions will be subjected to change without notice due to the 
improvement of performance or functions. 
 
   Contact our sales representative or agent if any page of the operating instructions is lost or page order is 
incorrect. 
 
   Contact our sales representative or agent if any point in this operating instructions is unclear or if there 
are any inaccuracies. 
 
   No part of this operating instructions may be reproduced in any form without the expressed written 
permission of our company. 
 
 

CAUTION

 

Our company guarantees the product under certain warranty conditions.  Our company in no 
way shall be responsible for any loss of content or damage of content. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Содержание MDF-U5312

Страница 1: ...Series Please read these instructions carefully before using this product and save this operating instructions for future use See page 33 for all Model numbers MDF U5312 Biomedical Freezer Operating Instructions ...

Страница 2: ...PERATURE SETTING P 15 SETTING OF ALARM RESUME TIME P 16 REMOTE ALARM TERMINAL P 17 ALARMS SAFETY FUNCTIONS P 18 ROUTINE MAINTENANCE P 19 Cleaning of cabinet P 19 Defrosting P 19 TROUBLESHOOTING P 20 DISPOSAL OF UNIT P 21 Recycle of battery P 21 Decontamination of unit P 21 DISPOSAL OF BATTERY P 26 TEMPERATURE RECORDER OPTION P 27 Setting of MTR 4015LH P 27 Setting of MTR G85A or MTR G85C P 29 SPEC...

Страница 3: ...uctions will be subjected to change without notice due to the improvement of performance or functions Contact our sales representative or agent if any page of the operating instructions is lost or page order is incorrect Contact our sales representative or agent if any point in this operating instructions is unclear or if there are any inaccuracies No part of this operating instructions may be rep...

Страница 4: ...s could result in injury to personnel and damage to the unit and associated property Symbol shows this symbol means caution this symbol means an action is prohibited this symbol means an instruction must be followed Be sure to keep this operating instructions in a place accessible to users of this unit Some warning and or caution labels are attached on the unit Following shows the description of s...

Страница 5: ...ower supply outlet is not grounded it will be necessary to install a ground by qualified engineers Never ground the unit through a gas pipe water main telephone line or lightning rod Such grounding may cause electric shock in the case of an incomplete circuit Connect the unit to a power source as indicated on the rating label attached to the unit Use of any other voltage or frequency other than th...

Страница 6: ...unction Disconnect the power supply plug if there is something wrong with the unit Continued abnormal operation may cause electric shock or fire When removing the plug from the power supply outlet grip the power supply plug not the cord Pulling the cord may result in electric shock or fire by short circuit Disconnect the power supply plug before moving the unit Take care not to damage the power co...

Страница 7: ...y cause a heat or ignition Never store corrosive substances such as acid or alkali in this unit if the container cannot be sealed These may cause corrosion of inner components or electric parts Check the setting when starting up of operation after power failure or turning off of power switch The stored items may be damaged due to the change of setting Be careful not to tip over the unit during mov...

Страница 8: ... 40 o C Maximum relative humidity 80 for temperature up to 31 o C decreasing linearly to 50 relative humidity at 40 o C Mains supply voltage fluctuations up to 10 of the nominal voltage Transient overvoltages up to the levels of OVERVOLTAGE CATEGORY II Temporary OVERVOLTAGES occurring on the mains supply Applicable pollution degree of the intended environment POLUTION DEGREE 2 in most cases ...

Страница 9: ...8 FREEZER COMPONENTS 12 Back side 11 14 9 8 15 4 7 5 6 10 13 13 1 2 3 ...

Страница 10: ... the outer door turn this latch downward A padlock is also available 9 Storage container Made of styrol resin Be careful not to damage the container by a scraper at the time of defrosting 10 Control panel To display the temperature setting and running condition See page 10 for details 11 Remote alarm terminal Used to notify an alarm condition of the unit to remote location See page 17 for details ...

Страница 11: ...shed Pressing this key again after setting desired temperature the setting is stored into computer memory If there is no key operation for 90 seconds during the setting mode the setting mode is invalid automatically See page 14 for the details 6 Defrost key DEF By pressing this key for 5 seconds the refrigerating operation is stopped Pressing this key again after defrosting leads resumption of the...

Страница 12: ... cause vibration or noise WARNING Install the unit on a sturdy floor If the floor is not strong enough or the installation site is not adequate this may result in injury from the unit falling or tipping over Select a level and sturdy floor for installation This precaution will prevent the unit from tipping Improper installation may result in water spillage or injury from the unit tipping over A lo...

Страница 13: ...g feet Extend the leveling feet by rotating them counterclockwise to contact them to the floor Ensure the unit is level Fig 1 3 Fixing the unit Two fixtures are attached to the rear of the frame Fix the frame to the wall with these fixtures and rope or chain Fig 2 4 Ground earth The ground earth is for preventing the electric shock in the case of the electrical insulation is somehow degraded Alway...

Страница 14: ...g the operation of the freezer for the first time the alarm lamp ALARM lights after the start of operation When the chamber temperature reaches around the set temperature then the alarm lamp goes out The remote alarm is not activated If the battery switch is turned on before turning on the power of the freezer the temperature alarm is activated and the buzzer sounds and the remote alarm is also ac...

Страница 15: ...essing set key SET is not memorized Although the value of the chamber temperature setting can range from 18 o C to 35 o C the guaranteed temperature with no load is 30 o C when the ambient temperature is 35 o C Key lock function This unit is provided with the key lock function When the key lock is ON change of temperature setting through the key pad is not possible The key lock is set to OFF at th...

Страница 16: ...ue shift key and scroll the figure to 1 The first digit is flashed 4 Press set key SET The first digit is flashed 5 Scroll the figure to 005 by using digit shift key and numerical value shift key When pressed the figure of settable digit changes When pressed the changeable digit moves 6 Press set key SET Alarm temperature is memorized and the current chamber temperature is displayed The set mode r...

Страница 17: ...ss digit shift key for 5 seconds The first digit is flashed 3 Set the figure to F25 with the digit shift key and numerical value shift key The settable digit is shifted When pressed the figure of settable digit changes 4 Press set key SET The current resume time is displayed The second digit is flashed 5 Set the figure to 020 with the numerical value shift key When pressed the figure of settable d...

Страница 18: ... terminal Contact capacity is DC 30 V 2 A Contact output between COM and N O between COM and N C At normal Open Close At abnormal Close Open Note The remote alarm is stopped by pressing buzzer stop key BUZZER as the remote alarm is operated in conjunction with the buzzer except for the power failure alarm The remote alarm terminal is in alarm status when the power supply plug is unplugged because ...

Страница 19: ...te alarm Temperature control sensor abnormality If the sensor is disconnected Alarm lamp is flashed E11 and chamber temp are displayed alternately Intermittent tone Remote alarm Normal operation If the sensor is short circuited Alarm lamp is flashed E12 and chamber temp are displayed alternately Battery check When about 3 years has passed with power switch ON _F01 and chamber temp are displayed al...

Страница 20: ...ct is refrigerated by the direct cooling When it uses for a long time frost appears on the chamber wall side It cannot be cooled down when there is much amount of frost The defrost is mentioned in the following Use the scraper provided for removing the frost if the freezer operation must be continued Pay attention not to impact or damage the inner wall Forced defrost 1 When defrosting temporarily ...

Страница 21: ...perature is not inputted properly The freezer is in the direct sunlight There is any heating source near the freezer A rubber cap and insulation for the access port are not set correctly You put too many unfrozen articles into the freezer compartment Noise The freezer is not installed on the sturdy floor The freezer is not leveled with the leveling feet There is anything touching the frame The fre...

Страница 22: ...emove doors to prevent accidents such as suffocation Recycle of battery The unit contains a rechargeable battery The battery is recyclable At the end of it s useful life check with you local solid officials option or proper disposal Label indication is obliged to comply with Taiwanese battery regulation Decontamination of unit Before disposing a biomedical freezer with biohazardous danger decontam...

Страница 23: ...spose of them correctly at your local community waste collection recycling centre Please help us to conserve the environment we live in German Für EU Staaten Das Symbol und das erwähnte Wiederverwertungssystem gelten nur für die Länder der EU und nicht für andere Länder oder Gebiete in der Welt Die Produkte von Panasonic werden aus hochwertigen Materialien und Komponenten gefertigt die sich wieder...

Страница 24: ...u éliminées de manière inappropriée peuvent s avérer potentiellement dangereuses pour la santé humaine et pour l environnement Cependant ces matières sont nécessaires au bon fonctionnement de votre appareil ou de votre machine Pour cette raison il vous est demandé de ne pas vous débarrasser de votre appareil ou machine usagé avec vos ordures ménagères Spanish PARA USUARIOS DE LA UNION EUROPEA El s...

Страница 25: ...em recolha de lixo local Por favor ajude nos a conservar o ambiente Italian PER UTENTI UE Il simbolo e i sistemi di riciclaggio descritti di seguito si applicano esclusivamente ai paesi dell UE Questo prodotto Panasonic è stato progettato e realizzato con materiali e componenti di elevata qualità che possono essere riciclati e o riutilizzati Il simbolo di riciclaggio mostrato di seguito indica che...

Страница 26: ...bij uw plaatselijk afvalinzameling recyclingcentrum in uw buurt Help ons het milieu waarin wij leven in stand te houden Swedish FÖR ANVÄNDARE INOM EU Den symbolmärkning och de återvinningssystem som beskrivs här nedan gäller länder inom EU och gäller inte länder i någon annan del av världen Din Panasonic produkt har konstruerats och tillverkats med delar och material av hög kvalitet som kan återvi...

Страница 27: ...off the power switch and disconnect the power supply plug 2 Remove 6 screws fixing the top cover with a screw driver and remove the top cover Fig 1 3 Disconnect the battery connector and remove 2 screws fixing the battery mounting plate Fig 3 4 Take out the battery Fig 4 5 Follow the procedure for recycling or proper disposal Handling of battery Cover the battery terminal with an insulating tape t...

Страница 28: ... 4015LH Pull the knob on the upper part of the temperature recorder forward to change the recording paper or battery Setting of recording paper 1 The information on the temperature recording paper is shown in Fig 1 2 Pull the cartridge up after opening the top lid The lid can be opened by turning the knob counterclockwise See Fig 2 3 As shown in Fig 3 insert the recording paper with the begin tab ...

Страница 29: ...rding paper has been fed into the used recording paper compartment send all the recording paper in the compartment first turning the gear Battery replacement To replace the battery turn the knob counterclockwise to open the lid Place the battery in the battery case according to the plus minus indications on the bottom of the battery case See Fig 6 Note This temperature recorder is designed for the...

Страница 30: ...ress both sides of the ink pen arm as indicated by the arrows to open the head clamp at A and B See to Fig 2 illustration 1 4 Position the ink pen so that the guide pins fit into the guide holes on the ink pen arm See to Fig 2 illustration 2 5 Press the two sides of the head clamp as indicated by the arrows to secure the ink pen See to Fig 2 illustration 3 From the side view the ink pen should fit...

Страница 31: ...2 days The center of the recording paper is divided into 32 equal sections The lines extending from these lines serve as the 32 day time scale Fig 4 Stopping recording 1 Turn OFF the power switch 2 When recording is stopped for a prescribed period place the caps back on the ink pen to prevent the ink from evaporating Replacing the recording paper 1 Slightly raise the end of the pen lifter and remo...

Страница 32: ...raper 6 small baskets for upper chamber W290 x D536 x H136 mm 4 large baskets for lower chamber W290 x D536 x H238 mm Weight 134 kg Battery For power failure alarm Nickel metal hydride battery DC 6 V 1100 mAh Automatic charge 5HR AAC Optional component Temperature recorder MTR 4015LH Temperature recorder MTR G85A MTR G85C Recorder fixing for temperature recorder MPR S30 MPR S7 Interface board MTR ...

Страница 33: ...t temperature 35 o C no load Temperature control range 20 o C to 30 o C Power voltage AC 110V AC 115 V AC 220 V AC 220 V AC 230 V 240 V Rated frequency 60 Hz 60 Hz 50 Hz 60 Hz 50 Hz Rated power consumption 265 W 275 W 240 W 285 W 255 W 290 W Noise level 40 dB A background noise 20 dB Maximum pressure 1 80 MPa Note The unit with CE mark complies with EC directives ...

Страница 34: ...the unit Unit interior No contamination Decontaminated Contaminated Others 3 Instructions for safe repair maintenance of the unit a The unit is safe to work on b There is some danger see below Procedure to be adhered to in order to reduce safety risk indicated in b below Date Signature Address Division Telephone Product name Model Serial number Date of installation Biomedical Freezer MDF Please de...

Страница 35: ...Printed in Japan 7FB6P151607001 S0412 10412 1 1 1 Sakata Oizumi Machi Ora Gun Gunma 370 0596 Japan Panasonic Healthcare Co Ltd 2012 ...