33
TEMPERATURE RECORDER (OPTION)
Installation of MTR-G85A/MTR-G85C
1.
Remove 4 screws (A, B, C, D) on the side of the
front panel to take off the front panel. Then remove 4
screws on the cover for the recorder mounting space
and take off the cover (Fig. 1).
2.
Fix the temperature recorder to the front panel with
fixture and screws enclosed with the recorder.
Note
:
Before fixing the temperature recorder, connect the
recorder connector with the connector on the right
back side of the control panel (Fig. 2)
3.
Lead the recorder sensor to the chamber through
the access port at the bottom of the chamber.
Note:
The port is covered with a thermal insulation and
rubber cap. Remove those covers before passing
the recorder sensor.
4.
Attach the sensor cover to the right side of the
chamber with 2 screws on the chamber wall.
Fig. 1
Front panel
A
B
C
D
Cover
< Back of the Front panel >
Fig. 2
Connector
Temperature
recorder
Fixture