16
ALARM TEMPERATURE SETTING
Table 4 Procedure for setting low temperature alarm
Description of operation
Key operated
Indication after operation
1
----
The current chamber temperature is
displayed.
2
Press numerical value shift key for
5 seconds.
The right digit is flashed.
3
Press numerical value shift key and
scroll the figure to 2.
When pressed, the figure of settable
digit increases.
4 Press set key.
SET
The right digit is flashed.
5
Set the temperature to -05 with the
digit key and numerical value shift
key.
Pressing the key shifts the digit which can be set.
When pressed, the figure of settable
digit increases.
6 Press set key.
SET
Alarm temperature is memorized and
the current chamber temperature is
displayed.