APPENDIX
English
18
Q
The
mark does not light when an arm is moved.
A
The
mark is a function to notify you that your arm is bent or to indicate extra loads on
the cuff.
Therefore, it may not indicate motions (bending an arm slightly, etc.) that do not affect the cuff.
If you are concerned about the result, retry the measurement.
Troubleshooting
Display
Status before error
Cause and solution
is displayed.
The indication disappears in the
middle of the operation. (the
mark is displayed.)
Batteries are exhausted. (See page 6.)
is displayed.
Your arm or body moved.
Has the unit detected movement? (See
page11.)
Pressurization to more than
280 mm
Hg
was performed. (Cuff
pressurizes several times.)
Were measurements taken according to
the proper procedure and in the correct
posture? (See page9.)
Pressure decreased suddenly.
Is the cuff wrapped properly?
●
Is the cuff plug inserted correctly into
●
the socket?
(See page8.)
Cuff does not inflate.
The measurement is not possible
though the cuff inflates.
Cuff pressurizes several times.
is displayed.
Failure of the main unit
Take it to the store where it was purchased
for testing and repairs.