Specifications
8
VeriDri™ User’s Manual
•
Max. Loop Resistance:
R = 50
(PSV–7), where PSV = Power Supply Voltage
Example:
For a PSV = 24 VDC, R = 50
(24–7) = 850
•
Cable:
2 m, standard (consult factory for custom lengths)
7.3
Mechanical Specifications
Sample Connection
3/4-16 straight, male thread with o-ring
Operating Pressure
5
m Hg to 5,000 psig (345 bar)
Enclosure
Type 4X/IP67
European Compliance
See the
EC Declaration of Conformity
at the end of this manual
Dimensions
•
Overall:
8.51
1.125 in. (21.62
2.86 cm)
•
Electronics Module w/cable:
5.93
1.125 in. (15.06
2.86 cm)
•
Weight:
5 oz (140 g)
7.4
Moisture Sensor Specifications
Sensor Type
Thin-film aluminum oxide moisture sensor probe
Calibration
Each sensor is individually computer-calibrated against known moisture concentrations, traceable to NIST
Calibration Interval
Factory sensor recalibration by Panametrics is recommended every
6-12 months, depending on the application
Calibration Data
Factory-calibrated, stored in
FLASH
memory
Flow Rate
•
Gases:
Static to 10,000 cm/s linear velocity, at a pressure of 1 atm
•
Liquids:
Static to 10 cm/s linear velocity, at a density of 1 g/cc
3/4-16 UNF-2A
2.58 [65.53]
2.58 [65.53]
1.125 [28.58]
3.35 [85.10]
MIN
CABLE BEND
ALLOWANCE
Содержание VeriDri
Страница 1: ...panametrics com 916 095 D Dec 2021 VeriDri Moisture Transmitter User s Manual...
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 8: ...Contents vi VeriDri User s Manual no content intended for this page...
Страница 18: ...Warranty 10 VeriDri User s Manual no content intended for this page...
Страница 20: ...no content intended for this page...
Страница 21: ......