Inserting the Transmitter into the System
2
VeriDri™ User’s Manual
Figure 1: Sample System Example
3.
Inserting the Transmitter into the System
Complete the steps below to install the VeriDri:
1.
Make sure the sintered or sheet stainless-steel shield is in place. The shield protects the aluminum oxide sensor
from damaging elements in the process.
2.
The probe is mounted into the process via the 3/4-16 straight male thread located on the probe. Thread the
probe end of the transmitter into the process/sample system fitting, making sure not to cross thread it.
CAUTION!
If you are mounting the VeriDri directly into the process line, you must consult
Panametrics for proper installation instructions and precautions before beginning the
following procedure.
Sample
Cell
Sample
Inlet
Sample
Outlet
Vent
Shield
3/4-16 UNF-2A