 Cleaning the Transmitter Probe

6

VeriDri™ User’s Manual

6.2

Soaking the Sensor and Shield

While wearing protective gloves, complete the following steps:

1.

Place the sensor in the first container of hexane or toluene and allow it to soak for 10 minutes.

2.

Remove the sensor from the hexane or toluene and soak it in the container of distilled water for 10 minutes.

3.

Remove the sensor from the distilled water and soak it in the second (clean) container of hexane or toluene for 10 

minutes.

4.

Remove the sensor from the hexane or toluene and set it aside until the shield has completed its cleaning cycle.

5.

Clean the shield by repeating steps 1 to 3 above. To ensure the removal of any contaminants that may have 

become embedded in the porous walls of the shield, swirl the shield in the solvents during the soaking 

procedures.

6.

Remove the shield from the hexane or toluene.

7.

Carefully replace the shield over the exposed sensor, 

without touching the sensor

.

8.

Place the sensor with the installed shield in an oven set at 50°C ± 2°C (122°F ± 3.6°F) for 24 hours.

6.3

Evaluating the Probe

To evaluate the probe, complete the following steps:

1.

Re-connect the cable to the transmitter module and measure the ambient dew point. Make sure you measure 

the same ambient air as measured in step 2 on 

page 5

.

2.

Compare the two ambient air readings. If the new ambient air reading is within ±2°C (±3.6°F) of the first reading, 

the cleaned probe is properly calibrated and may be reinstalled. If the two readings not within that tolerance, 

proceed to step 3 below.

3.

If the probe is still not reading the ambient air accurately, repeat the cleaning procedure using soaking times 

that are 5 times the previous cleaning sequence, until two consecutive ambient air readings are within ±2°C 

(±3.6°F).

If the above cleaning procedure does not result in accurate readings, contact Panametrics for assistance.

CAUTION!

Any attempt to open the module or remove the probe from the module will void your 

warranty.

CAUTION!

Do not place the transmitter module into the solvents. You should only insert the sensor 

portion of the transmitter into the solvents. Also, do not allow the sensor to come into 

contact with the surfaces of the cleaning containers or any other hard surface.

Shield

Содержание VeriDri

Страница 1: ...panametrics com 916 095 D Dec 2021 VeriDri Moisture Transmitter User s Manual...

Страница 2: ......

Страница 3: ...al contains one or more registered trademarks of Baker Hughes Company and its subsidiaries in one or more countries All third party product and company names are trademarks of their respective holders...

Страница 4: ...ii no content intended for this page...

Страница 5: ...to the auxiliary equipment Examples include safety glasses protective headgear safety shoes etc Unauthorized Operation Make sure that unauthorized personnel cannot gain access to the operation of the...

Страница 6: ...vironment In order to avoid the dissemination of those substances in our environment and to diminish the pressure on the natural resources we encourage you to use the appropriate take back systems Tho...

Страница 7: ...onnections 3 5 Operating the Transmitter 4 5 1 Powering Up 5 5 2 Error Handling 5 6 Cleaning the Transmitter Probe 5 6 1 Removing the Transmitter 5 6 2 Soaking the Sensor and Shield 6 6 3 Evaluating t...

Страница 8: ...Contents vi VeriDri User s Manual no content intended for this page...

Страница 9: ...unit be installed in a sample system to protect the probe from coming into contact with damaging elements in the process Before constructing a sample system you should consult a Panametrics Applicati...

Страница 10: ...om damaging elements in the process 2 The probe is mounted into the process via the 3 4 16 straight male thread located on the probe Thread the probe end of the transmitter into the process sample sys...

Страница 11: ...secure the connectors together by sliding the metal sleeve on the cable over the connectors and turning it clockwise until it is tight 2 Using the flying leads at the other end of the transmitter cabl...

Страница 12: ...ly or worry about losing data during a power loss Probes may need to be cleaned occasionally depending on the application Consult a Panametrics Applications Engineer for recommended cleaning intervals...

Страница 13: ...latex gloves An oven set at 50 C 2 C 122 F 3 6 F for drying the probe A 1 1 8 in wrench Refer to the following sections to properly clean the probe 6 1 Removing the Transmitter Complete the following...

Страница 14: ...50 C 2 C 122 F 3 6 F for 24 hours 6 3 Evaluating the Probe To evaluate the probe complete the following steps 1 Re connect the cable to the transmitter module and measure the ambient dew point Make su...

Страница 15: ...erating Temperature 40 to 60 C 40 to 140 F Storage Temperature 70 C 158o F maximum Note The probe should be stored with the plastic cover and desiccant packet threaded onto the probe Store it in a coo...

Страница 16: ...onics Module w cable 5 93 1 125 in 15 06 2 86 cm Weight 5 oz 140 g 7 4 Moisture Sensor Specifications Sensor Type Thin film aluminum oxide moisture sensor probe Calibration Each sensor is individually...

Страница 17: ...pose and warranties arising from course of dealing or usage or trade Return Policy If a Panametrics instrument malfunctions within the warranty period the following procedure must be completed 1 Notif...

Страница 18: ...Warranty 10 VeriDri User s Manual no content intended for this page...

Страница 19: ...trial Locations EN 61326 2 3 2006 EN 60529 2001 IP67 following the provisions of the 2004 108 EC EMC Directive The unit listed above and any ancillary equipment supplied with it do not bear CE marking...

Страница 20: ...no content intended for this page...

Страница 21: ......

Страница 22: ...one or more countries All third party product and company names are trademarks of their respective holders 916 095 D Customer Support Centers U S A 1100 Technology Park Drive Billerica MA 01821 4111...

Отзывы: