Appendix A. CE Mark Compliance
Appendix A. CE Mark Compliance
A.1 Introduction
For CE Mark compliance, the Model XMT868i flow transmitter must be wired in accordance with the instructions in this
appendix.
IMPORTANT:
CE Mark compliance is required for all units intended for use in EU countries.
A.2 Wiring
The Model XMT868i must be wired with the recommended cable, and all connections must be properly shielded and
grounded. Refer to
below for the specific requirements.
Note:
If the Model XMT868i is wired as described in this appendix, the unit will comply with the EMC Directive.
Table 21: Wiring Requirements
Connection
Cable Type
Ground Termination
Transducer Armored RG62 a/U
Grounded using a cable gland.
Input/Output Armored 22 AWG shielded (e.g. Baystate #78-1197) with
armored material added to outside of jacket
Grounded using a cable gland.
Power
Armored 14 AWG 3 conductor
Grounded using a cable gland.