DigitalFlow™ XMT868i Startup Guide
5
Chapter 4. Specifications
4.3 Transducer
4.3.1
Wetted Transducers
Material:
Standard:
316 Stainless Steel.
Optional (for isolating PanAdapta Plugs
):
Hastelloy, Monel, Titanium, Duplex, CPVC, PVDF and others.
Process Connection:
Standard:
1-in. NPTM, 3/8-in. NPTM.
Optional:
RF flange, socket weld, fuse bond and others.
Maximum Pressure:
Standard:
0 to 3,000 psig (0.1013 to 20 MPa).
Optional:
Higher pressures on request.
Temperature Range:
Standard:
–40° to 212°F (–40° to 100°C).
Optional (overall range):
-310
o
to 932
o
F (-190 to 500
o
C).
Mounting:
Spoolpiece, hot tap, or cold tap.
Area Classifications:
Standard:
General-purpose
Optional:
Explosion proof:
Class I, Div. 1, Groups B, C & D; Class II, Groups E, F and G; Class III
Flameproof:
II 2 G EEx d IIC T6
Weatherproof:
Type 4/IP65.
Submersible.
Note:
Wetted transducers for special applications and flowcells are also available. Consult the factory for details.
Содержание DigitalFlow XMT868i
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 28: ...Chapter 1 Installation 20 DigitalFlow XMT868i Startup Guide...
Страница 34: ...Chapter 1 Installation DigitalFlow XMT868i Startup Guide 26...
Страница 52: ...Chapter 2 Initial Setup 18 DigitalFlow XMT868i Startup Guide...
Страница 76: ...Appendix A CE Mark Compliance 10 DigitalFlow XMT868i Startup Guide...
Страница 86: ...Chapter 20 DigitalFlow XMT868i Startup Guide no content intended for this page...
Страница 90: ...2 DigitalFlow XMT868i Startup Guide...
Страница 92: ...2 Doc Tittle Manual Type...
Страница 93: ......