Chapter 4. Parts Replacement
1.
If necessary, replace the angle bracket on the main board, insert the screw (with flat washer and lock washer),
and screw it into the standoff on the main board. See
Illustration 11
in
IMPORTANT:
The small angle bracket on the terminal board must be positioned on the outside of the main board
with the previously removed nylon washer between them.
2.
Face the
User Program
side of the main board toward the rear of the power supply board. Orient the boards so
that the pins along the bottom of the main board align with connectors J2 and J6 on the power supply board,
and push the two boards firmly together.
Note:
If the thermal gap filler pad of the power supply board has been damaged, you must remove and replace it
before reinserting the power supply board. First, remove and discard the damaged pad. Then remove the
blue protective liner from the new pad and apply it with the white face against the power supply board.
3.
Secure the power supply board and heat sink to the main board by reinstalling the screws, flat washers and lock
washer previously removed from the corners of the power supply board. See
Illustration 11
in
.
4.
Refer back to
Illustration 7
in
. Carefully align the terminal board onto the main board/power
supply assembly and the two option boards (Slot 1 and Slot 2, where applicable), so that the connectors listed in
Table 5
below are facing each other. Then, firmly press the terminal board and the interconnect board toward
each other until all of the connectors are completely seated.
IMPORTANT:
With two option cards installed, it may require some patience to get all of the connectors properly
aligned. Do not force the assembly together, or some of the pins may not enter the sockets properly.
5.
Secure the assembly together by installing the three mounting screws, flat washers and lock washers previously
removed from the terminal board. Be sure to install one end of the green grounding jumper wire under the screw
indicated in
Illustration 7
in
. The smaller screw should be used at the angle bracket location,
and secured with a nut, lock washer and flat washer.
Proceed to the next section for instructions on installing the reassembled circuit board assembly in the electronics
enclosure.
Table 5: Mating Connectors for Terminal Board
Terminal Board
Mating Board
1
J7 Connector (Large Three-Pin Connector Plug)
Power Tab on AC Power Supply Board
2
J6 Connector (Large Three-Pin Connector Plug)
Power Tab on DC Power Supply Board
3
P7 Connector
J7 Connector on Main Board
4
P10 Connector
J9 Connector on Main Board
5
P9 Connector
J10 Connector on Main Board
6
P8 Connector
J8 Connector on Main Board
7
P21 Connector
J1/J3 Connector on I/O Board, Slot 2 Option
8
P22 Connector
P2 Connector on I/O Board, Slot 1
Содержание DigitalFlow XGF868i
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 24: ...Chapter 2 Error Codes 16 DigitalFlow XGF868i Service Manual no content intended for this page...
Страница 40: ...Chapter 4 Parts Replacement DigitalFlow XGF868i Service Manual 32...
Страница 44: ...Appendix A Service Record 36 DigitalFlow XGF868i Service Manual no content intended for this page...
Страница 70: ...Warranty 62 DigitalFlow XGF868i Service Manual no content intended for this page...
Страница 71: ......