8/19/03
910-198E
Model 8500
Page 5-17
are not functioning properly can be easily identified. To exit the test,
press [ENTER].
5.5.3 Hardware Test
The Hardware Test has two purposes. First, basic information, such as
software and probe type, is displayed. This information appears on the
top of the display:
Second, there is a debugging option located on the top area of the
display. If this option is turned ON, a new screen appears on the
Measurement display supplying hardware data information. The
following definitions refer to each field in the graphic below:
• Count: Analog/Digital converter number
• INT Index: Numerical location on the look-up table
• BL: Base Line (ball OFF position) A/D number
• FS: Full Scale (ball ON position) A/D number