Mattress will not inflate or if every other cell is not inflating:

•  Disconnect the control unit and make sure air is flowing from the open ports. This is 

done by disconnecting the control unit from the mattress and physically feeling the air 

blowing out from one port and then the other. It may take several minutes to switch 

from one port to the other depending on the control unit’s cycle time. If the control unit 

is not blowing air out of the ports, call for service.

•  Check to make sure the control unit is on.
•  Check to make sure the control unit is properly connected to the mattress.
•  Check to make sure the red CPR release valve is closed.
•  Open mattress and inspect the internal connections to make sure each individual air 

cell is properly connected and not leaking. If a defect is discovered, call for repair or 

replacement of defective component.

•  Visually inspect each air cell to make sure there are no cuts, punctures or tears in the 

material with the exception of the low air loss holes designated by the white circles 

on the air cells. If a defect is discovered, call for repair or replacement of defective 

component.

•  Check to make sure there are no kinks or leaks in the internal and external hoses. If a 

defect is discovered, call for repair or replacement of defective component.

Mattress is soft and will not get firm or Low Pressure Alarm is sounding:

•  Check control unit to make sure it is on and functioning properly. If the control unit is 

not functioning, call for service.

•  Inspect the air filter to make sure it is clean. The filter is located on the back of the 

control unit near the upper left corner. The filter should be white in color and free of 

dust or other debris. If the filter is not clean, replace with a new filter.

•  Visually inspect each air cell to make sure there are no cuts, punctures or tears in the 

material with the exception of the low air loss holes designated by the white circles 

on the air cells. If a defect is discovered call for repair or replacement of defective 

component.

NOTE:

 

The pressure in the mattresses lowers when the weight of the resident is 

removed from the mattress. The alarm will sound until the pump re-establishes the 

pressure level. In this situation wait about 5 minutes and check it again.

Control unit is not working:

•  Check to make sure the control unit is plugged into a live electrical socket.
•  Make sure the control unit is turned on.
•  Check the external fuses to make sure the fuses are in good condition. 

Make sure the 

control unit is unplugged from the power source. 

Locate the fuses on the side 

of the control unit near the on/off switch. Unscrew the fuse cap and inspect the fuse 

to see if the filament, small wire inside of the glass tube, is broken. If the filament is 

broken, install new fuse and reinsert the fuse cap. Be sure the replacement fuse is of 

the same voltage and amperage as the previous fuse.

Troubleshooting Guide

Содержание AIR APL MATTRESS

Страница 1: ...irect Supply Equipment Furnishings Please read this entire guide carefully and keep it for future reference This guide will provide you with instructions warnings warranty information and other important information about your Panacea mattress Share this information with your housekeeping nursing and maintenance staff to help ensure the mattress system is cared for properly ...

Страница 2: ...3 Product Description 4 5 Product Specifications 6 Warnings 7 Directions for Use 8 9 Transportation CPR 10 Cleaning 11 Preventive Maintenance 12 Storage 12 Troubleshooting Guide 13 Limited Warranty 14 Customer Service 15 Table of Contents ...

Страница 3: ...des alternating pressure and limited low air loss therapy The inflation levels are designed to maintain low interface pressures throughout the mattress to redistribute peak interface pressures during the cycle and to further relieve peak pressure points as the cycle alternates The low air loss feature helps reduce maceration and further skin breakdown by keeping the skin surface cool and dry Defin...

Страница 4: ...in the event of a power failure CPR VALVE Allows for rapid deflation and is located at the head end of the mattress See directions for proper use Cover Top cover The antimicrobial top fabric is fire resistant fluid resistant tear resistant cleanable and the two way stretch cover is replaceable Bottom cover PVC vinyl bottom for durability and easy cleaning Mattress Height The total pressure managem...

Страница 5: ...owers the control unit on and off Firmness Dial Adjusts the pressure within the system Static button Allows for the alternating pressure functionality to be turned off Low Pressure Alarm Indicates loss of air pressure Alarm Silence button Silences the low pressure alarm ...

Страница 6: ... wrapped in moisture proof polyurethane material CONTROL UNIT 120V 60 Hz SURFACE SIZES 35 W x 8 H x 75 L 35 W x 8 H x 80 L 1 Resident s body cannot exceed the width of the mattress 2 Check the weight capacity of the bed to make sure both the mattress and bed are capable of supporting additional weight OPERATING MODES Panacea Air APL Alternating Pressure Provides 10 minute loading and unloading cyc...

Страница 7: ... unit immediately after using 2 Do not operate near water 3 Do not place or store product where it can fall or be pulled into a tub or sink 4 Do not place in or drop into water or other liquid 5 Do not reach for a product that has fallen into water Unplug immediately 6 Do not remove control unit cover Risk of electrical shock WARNING To reduce the risk of burns fire or injury to persons 1 Use this...

Страница 8: ...eep pocketed sheets are recommended not included Multiple layering of linens or under pads beneath the resident can negatively affect the mattress s pressure management capabilities and should be avoided unless recommended by a caregiver 8 Increase firmness if necessary for resident comfort support and immersion into the mattress for optimal pressure redistribution a A common method for sizing an ...

Страница 9: ...esident can negatively affect the mattress s pressure management capabilities and should be avoided unless recommended by a caregiver CAUTION Be careful not to puncture the mattress with needles or sharp instruments This may result in loss of integrity of the cover or internal air system NOTE This product is only one element of care in the prevention and treatment of pressure ulcers by medical pro...

Страница 10: ...ce WARNING DO NOT attempt to use the mattress as a carrying device as it is not designed to carry a resident s weight unsupported CPR The standards for life support recommended by the American Heart Association for performing CPR recommend a hard level surface for performing CPR moving the person to the floor if possible If unable to move the resident to another surface to perform CPR place a cras...

Страница 11: ...ce DO NOT heat steam autoclave or immerse the control unit in liquids 1 The following germicidal detergents disinfectants are recommended by the EPA as hospital disinfectants Hi Tor Germicidal Detergent Huntington Laboratories Inc Indiana EPA 303 91 2 Follow the preparation instructions provided with the germicidal detergent disinfectant solution to prepare the required amount of disinfectant solu...

Страница 12: ...uency for replacing depends on the air quality 2 The air filter is accessible from the outside of the pump it is approximately two inches by one inch covered by a plastic cover 3 As the filter is white the need to replace dirty filters is apparent because they will darken in color due to the collection of dirt Simply remove the plastic cover remove the filter and replace with a new filter 4 Replac...

Страница 13: ...rm or Low Pressure Alarm is sounding Check control unit to make sure it is on and functioning properly If the control unit is not functioning call for service Inspect the air filter to make sure it is clean The filter is located on the back of the control unit near the upper left corner The filter should be white in color and free of dust or other debris If the filter is not clean replace with a n...

Страница 14: ... limited warranty period identified below whichever is earlier Product Part Warranty Period Coverage Parts Anticipated Usable Device Life Electrical Components 12 months 3 years Mattress 2 years 5 years Cover 1 year 1 year Anticipated Usable Device Life is based on normal use with proper maintenance cleaning and storage You should still inspect monitor and care for the device as described in this ...

Страница 15: ...ng experience receive a quality mattress and enjoy outrageous customer service If you have any questions about the mattress you have purchased or would like to request warranty service please contact Direct Supply Equipment Furnishings 1 800 634 7328 6767 N Industrial Road Milwaukee WI 53223 SalesSupport DirectSupply com ...

Страница 16: ...635 N Industrial Road Milwaukee WI 53223 2017 Direct Supply Manufacturing Inc All Rights Reserved Direct Supply Panacea and all associated logos are the registered trademarks of Direct Supply Inc used under license 2387 Date of Printing November 2017 ...

Отзывы: