panacare B63 Скачать руководство пользователя страница 12





Guidance and manufacture’s declaration – electromagnetic emission

The device is intended for use in the electromagnetic environment specified below. The customer of the 
user of the device should assure that it is used in such an environment.

Guidance and manufacturer’s declaration - electromagnetic immunity

The device is intended for use in the electromagnetic environment specified below. The customer or the 
user of the device, should assure that it is used in such an environment.

Emission test

RF emissions
CISPR 11

Compliance

Electromagnetic environment – guidance

Group 1

The device use RF energy only for its internal function. 
Therefore, its RF emissions are very low and are not likely to 

cause any interference in nearby electronic equipment.

The device is suitable for use in all establishments, including 
domestic establishments other than domestic and those directly 
connected to the public low-voltage power supply network that 
supplies buildings used for domestic purposes.

RF emission
CISPR 11

Class B

Harmonic emissions
IEC 61000-3-2

Not applicable

Not applicable

Voltage fluctuations/ 
flicker emissions
IEC 61000-3-3

Recommended separation distances between 
portable and mobile RF communications equipment and the device 

The device is intended for use in an electromagnetic environment in which radiated RF disturbances are 
controlled. The customer or the user of the device can help prevent electromagnetic interference by 

maintaining a minimum distance between portable and mobile RF communications equipment 

(transmitters) and the device as recommended below, according to the maximum output power of the 

communications equipment.

Rated maximum output 
power of transmitter
(W)

Separation distance according to frequency of transmitter (m)

150 KHz to 80 MHz

80 MHz to 800 MHz

800 MHz to 2.7 GHz

For transmitters rated at a maximum output power not listed above, the recommended separation 

distance d in metres (m) can be estimated using the equation applicable to the frequency of the 

transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the 
transmitter manufacturer.

NOTE 1   At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE 2   These guidelines may not apply in all situations. Electromagnetic propagation is affected by 
absorption and reflection from structures, objects and people.

The MANUFACTURER should consider reducing the minimum separation distance, based on RISK
MANAGEMENT, and using higher IMMUNITY TEST LEVELS that are appropriate for the reduced
minimum separation distance. Minimum separation distances for higher IMMUNITY TEST LEVELS 
shall be

calculated using the following equation:

NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting 
antenna and the ME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is 
permitted by IEC 61000-4-3.

a ) For some services, only the uplink frequencies are included.

b ) The carrier shall be modulated using a 50% duty cycle square wave signal.

c ) As an alternative to FM modulation. 50% pulse modulation at 18 Hz may be used because while it 
does not represent actual modulation, it would be worst case.

Where P is the maximum power in W, d is the minimum separation distance in m, and E is the 
IMMUNITY
TEST LEVEL in V/m.

Test 

frequency

(MHz)

Band
(MHz)

Service

Modulation 

Maximum 
power 
(w)

Distance
(m)

IMMUNITY 
TEST LEVEL
(V/m)

385

450

710

745

780

810

870

930

430-470

GMRS 460,
FRS 460

FM
±5 kHz deviation
1 kHz sine

2

380-390

TETRA 400

Pulse Modulation
18 Hz

1.8

0.3

27

0.3

28

2

0.3

28

2

0.3

28

2

0.3

28

0.2

0.3

9

0.3

9

704-787

800-960

1700-1990

LTE Band 13,

17

GSM 800/900,
 TETRA 800,
 iDEN 820,
CDMA 850,
LTE Band 5

GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band 1,3
4,25;UMTS

Pulse
Modulation
18 Hz

Pulse
Modulation
217 Hz

Pulse
Modulation
217 Hz

0.2

1720

1845

1970

2450

2400-2570

5100-5800

Bluetooth,
WLAN
802.11 b/g/n,
RFID 2450,
LTE Band 7

WLAN 802.11
a/n

Pulse
Modulation
217 Hz

Pulse
Modulation
217 Hz

5240

5500

5785

a)

b)

b)

b)

b)

b)

b)

c)

a)

a)

Содержание B63

Страница 1: ...oad Guihua community Guanlan town Longhua new district 518100 Shenzhen PEOPLE S REPUBLIC OF CHINA Shanghai International Holding Corp GmbH Europe Eiffestrasse 80 20537 Hamburg Germany Panacare Officia...

Страница 2: ...ure and its Measurement 3 Components of Your Blood Pressure Monitor 4 Using Your Monitor for the First Time 5 Measurement Procedure 6 Care and Maintenance 7 Warranty service 8 Certifications 9 Technic...

Страница 3: ...this manual Warning The separate ac adapter which is intended to connect USB interface of Blood Pressure Monitor has not been evaluated according to IEC 60601 1 The safety of the product shall be reap...

Страница 4: ...microwave ovens during use These can lead to erratic results Note Do not attempt to service or repair this device yourself Should a malfunction occur refer to local distributor or the manufacturer Cau...

Страница 5: ...f 3 Average value function 5 Built in chargeable lithium battery 7 Cuff and main unit are integrated as one part 2 One user 120 sets memory 4 Irregular heartbeat checking 6 Low battery display 1 Heart...

Страница 6: ...nt Press the START STOP button again the measurement will stop b Assessment of the memory record Press MEM or MEM button once it will seek for the next previous assessment record c Delete record When...

Страница 7: ...xamination but serves to detect pulse irregularities at an early stage 5 4 Irregular Heartbeat Detector 5 3 Measure Procedure Picture 06 Refer to picture 06 The monitor is designed to take measurement...

Страница 8: ...ed Abnormal measurement results The cuff is too loose Pressure fluctuation is large when measuring The pressure exceeds the upper limit Abnormal or uncalibrated calibration data Authorized Representat...

Страница 9: ...ff with bladder must not be washed in a dishwasher clothes washer or submerged in water c Do not drop the monitor or treat it roughly in any way Avoid strong vibrations d Never open the monitor This i...

Страница 10: ...c fields near the unit This may result in incorrect operation of the unit 3 Caution This unit has been thoroughly tested and inspected to assure proper performance and operation 4 Caution this machine...

Страница 11: ...ironment NOTE UT is the a c mains voltage prior to application of the test level Guidance and manufacture s declaration electromagnetic immunity Conducted RF IEC 61000 4 6 Radiated RF IEC 61000 4 3 10...

Страница 12: ...ce d in metres m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitte...

Страница 13: ...nde Guihua Stadt Guanlan neuer Bezirk Longhua 518100 Shenzhen VOLKSREPUBLIK CHINA Shanghai International Holding Corp GmbH Europa Eiffestra e 80 20537 Hamburg Deutschland Panacare Official outlook com...

Страница 14: ...formationen zum Blutdruck und dessen Messung 3 Komponenten des Blutdruckmessger ts 4 Monitor zum ersten Mal verwenden 5 Messverfahren 6 Pflege und Wartung 7 Garantie Service 8 Zertifizierungen 9 Techn...

Страница 15: ...tur oder Feuchtigkeitsbedingungen au erhalb der im Abschnitt Spezifikationen dieses Handbuchs angegebenen Grenzen betrieben oder gelagert wird Achtung Das separate Netzteil zum Anschluss der USB Schni...

Страница 16: ...t latexfrei Vorsicht Selbstmessung bedeutet Kontrolle nicht Diagnose oder Behandlung Ungew hnliche Werte m ssen immer mit einem Arzt besprochen werden Unter keinen Umst nden sollten Sie die Dosierung...

Страница 17: ...lwertfunktion 4 berpr fung des unregelm igen Herzschlags 5 Eingebaute aufladbare Lithiumbatterie 6 Anzeige f r schwache Batterie 7 Manschette und Haupteinheit sind als ein Teil integriert 1 Heartbeat...

Страница 18: ...g wird gestoppt b Speicherdaten Dr cken Sie die Taste MEM oder MEM einmal um nach dem n chsten vorherigen Bewertungsdatensatz zu suchen c Datensatz l schen Wenn Sie nach Speicherdatens tzen suchen Hal...

Страница 19: ...hrt und die Messwerte im Speicher speichert 1 Setzen Sie sich bequem auf einen Stuhl mit den F en flach auf dem Boden 2 Dr cken Sie die START STOP Taste um die Messung zu starten Strecken Sie Ihren A...

Страница 20: ...zschlag erkannt oder Blutdruck berechnet Abnormale Messergebnisse Die Manschette ist zu locker Die Druckschwankung ist beim Messen gro Der Druck berschreitet den oberen Messbereich Abnormal oder unkal...

Страница 21: ...ntfernt werden Die Manschette mit Blase darf nicht in einer Sp lmaschine oder Waschmaschine gewaschen oder in Wasser getaucht werden c Lassen Sie den Monitor nicht fallen und behandeln Sie ihn in kein...

Страница 22: ...ationen Modell B63 Gewicht 240 2 g Kein Netzteil im Lieferumfang enthalten Anzeige 30 58mm 2 28 x1 18 LCD Digitalanzeige Gr e 120 B x 66 L x 30 H mm 4 72 B x2 6 L x1 18 H Zubeh r 1 Hauptger t 1 Benutz...

Отзывы: