Note, however, that the ppm output is not disabled if no ppm conversion constants
have been entered via the Configuration software (page 23, section 5.5).
If the Physical Quantity (see software, page 20, section 5.1) is set to ppm, but no
constants have been entered, the relay operation and appearance of REL legends
in the display will be unpredictable and meaningless.
Caution: The measured variable ppm (x) should only be used if
C1 to C4 constants have been loaded.
70
62
NC
NC
11
11
C
C
12
12
NO
NO
1
3
1
3
[ppm, %,°C/°F]
1
2
3
1
2
3
8°C
70
62
NC
NC
14
14
C
C
15
15
NO
NO
16
16
[ppm, %,°C/°F]
4
5
6
4
5
6
8°C
Switching relay 1:
Switching relay 2:
If relay 1 has tripped (ON),
then REL1 is displayed.
If relay 2 has tripped (ON),
then REL2 is displayed.
Hardware
13
i n s t r u c t i o n
b o o k l e t
Pall WS08 Series
Water Sensor
ON
Relay status
Switching off
Switching on
Switching point
Hysteresis
OFF
ON
Relay status
Switching off
Switching on
Switching point
Hysteresis
OFF
AB