Document #: 020-17475-00 Revision: E
Page #: 48 of 63
Press the setup button on the Main menu screen
Select the “Date/Time” function using the up and down arrows
Press the curved arrow to go to the next screen
Press the button with the date, adjust the date to current, then press
Enter
Press button with the time, adjust time to current, then press
Enter
Press
to return to the Settings list
3.
Change password for Supervisor and Operator levels
Login to the system as Supervisor
Press the setup button on the Main menu screen
Select the “Password” function using the up and down arrows
Choose the level of access to which the change of password is required
Press the curved arrow to go to the next screen
Press the curved arrow to go to the entry screen
Enter the new password twice as prompted and press
Change
Press
to return to the Settings list
4.
Adjust auto logout time
Login to the system as Supervisor
Press the setup button on the Main menu screen
Select the “Auto Logout” function using the up and down arrows
Press the curved arrow to go to the entry screen
Enter the time for auto logout then press
Enter
5.
View runtime counter indicators for System runtime and Motor runtime
Login to the system as Supervisor
Press the setup button on the Main menu screen
Select the “Runtime Hrs” function using the up and down arrows
Press the curved arrow to go to the view screen
The screen will indicate accumulated runtime separately for both system
and motor rotation
Press
to return to the Settings list
6.
Export of Recipe library to memory card
Insert a SD card into the Memory slot (See section O)
Login to the system as Supervisor
Press the setup button on the Main menu screen
Select the “Recipe Imp/Exp” function using the up and down arrows
Press the curved arrow to open the next screen (Figure 58)
Press the
File number
button to go to the entry screen
On the entry screen input the file number (0-9) you wish to export and
press
Enter
Press the
Export
recipes
button and confirm when prompted (Figures 59
and 60).
The entire library will be exported to the Micro SD card under filename
LIBRARYX.BAK, where X is the file number entered on the previous step.
A confirmation notice will appear after the file writing is finished (this may
take several minutes)
Acknowledge the notice and press
to return to the Settings list
Содержание LevMixer DB-300
Страница 53: ...Document 020 17475 00 Revision E Page 53 of 63 Figure 62 Levmixer components bottom view ...
Страница 55: ...Document 020 17475 00 Revision E Page 55 of 63 Figure 64 Electrical Schematic Cage ...
Страница 62: ......