Document #: 4100022 Revision: F
Page #: 16 of 21
11. DISCONNECTING DRIVE UNIT FROM TANK
12.1
When mixing is complete, turn the RPM Regulator to zero (0) and turn the drive
switch to the STOP position. (Ensure that electrical power remains connected to the
machine when mixing multiple bags consecutively.)
12.2
Firmly hold the drive unit handle and raise the drive unit slightly to release the locking
lever. Release the locking lever by pulling it toward the control box.
12.3
Carefully lower the rear wheels of the drive unit to the floor and roll the drive unit on
its rear wheels away from the dolly.
12.4
Press down on the drive unit handle just until the guide bearings are free from the
guide rails.
12.5
Pull the drive unit slightly farther away from the dolly and carefully lower the front
wheel to the ground.
12.6
The tank can now be wheeled to another station on the elevated dolly. The
Superconducting Drive Unit remains ready for mixing.
12.7
To drain fluid from the bag, unclamp the bottom drain tube.
12.8
When the bag is completely drained, remove the bag by carefully pulling the impeller
seat and the drain tube out of their respective ports.
12.9
Remove the centering aligner from the tank and return to the supplied Accessories
Box for future use. The Interface should remain in its locked position for future use.
12.10
Always put the protective shell back on the bag before disposal.
12.11
Dispose of plastic bag.
Содержание DU005-US
Страница 6: ...Document 4100022 Revision F Page 6 of 21 4 ELECTRICAL SCHEMATIC 115 VAC ...
Страница 7: ...Document 4100022 Revision F Page 7 of 21 230 VAC ...
Страница 18: ...Document 4100022 Revision F Page 18 of 21 ...
Страница 20: ......