Micro 800 Optical DO Instruction Manual
Rev 1
© 2014 Palintest Ltd.
www.palintest.com
Page 36 of 41
Troubleshooting Continued …
Problem
Cause / Remedy
Dissolved Oxygen
readings are inaccurate
or unstable.
9
The DO sensor may need calibrating. Recalibrate.
9
The DO sensor cap may be fouled. Clean the DO Cap.
9
The DO Cap may be damaged. Inspect and replace if
necessary.
9
Calibration may have been carried out at an extreme
temperature. Recalibrate at a temperature as close to the
sample temperature as possible.
EC readings are
inaccurate or unstable.
Error 09 shows during
calibration of EC.
9
Have you got the Probe Sleeve fitted? EC will not work without
the Probe Sleeve fitted. The only exception to this rule is if the
Probe is being used with a Flowcell.
9
The Probe may not be inserted deep enough into the sample
being measured. Ensure the sample level reaches the top set
of holes on the Probe sleeve.
9
Trapped air bubbles may be causing problems. Tap and swish
the Probe to dislodge them.
9
The Probe Sleeve may be loose. The Probe Sleeve must be
absolutely rigid with respect to the Probe Body for correct EC
operation.
9
The EC electrode may need recalibrating. Recalibrate.
9
The EC electrode may be dirty. Clean the EC electrode then
recalibrate.