 The MultiStim ECO and its listed accessories must be disposed of in compliance 

with national provisions. Return the old device to the corresponding registered 

EAR (German register for waste electric equipment) collection container or 

return it to the manufacturer.

Technical description

The MultiStim ECO generates reproducible rectangular pulses with a frequency 

of 1 Hz and incrementally adjustable stimulation current. The setting range for 

the pulse current is 0.2 to 2.0 mA for stimulation cannulas and catheters.
If a current pulse is delivered and the patient circuit closed, the green LED next 

to the stimulation button blinks for visual control; for acoustic control, a short 

signal tone synchronous to the stimulation rhythm can be heard, i.e. stimula-

tion current flows through the patient.
If the green LED next to the stimulation button is on, either the patient circuit 

is not closed or there is a resistance > 12 kΩ; in these cases, there is no acoustic 

signal, i.e. no stimulation current or a stimulation current lower than selected 

one flows through the patient. This is additionally indicated by the yellow LED 

next to the unequal sign (≠).

Technical data

Type

MultiStim ECO

Degree of protection against 

electric shock EN 60601-1

BF

Protection class IEC 60529

IP54

Battery

2 x 1.5V LR1 (N size, LADY battery) 

Stimulation current

Max. 2 mA

Resistance range:

0 Ω - 12 kΩ

Stimulation voltage 

Max. 24 V

Stimulation frequency

1 Hz

Stimulation pulse width

0.1 ms

Operating conditions

Temperature:   10 °C to 30 °C
Humidity:  

20 % – 65 %

Pressure range:   700 hPa to 1060 hPa

Transport and storage condi-

tions 

Temperature:   10 °C to 30 °C
Humidity:  

20 % – 65 %

Pressure range:   700 hPa to 1060 hPa

Содержание MultiStim ECO

Страница 1: ...Nerve stimulation MultiStim ECO...

Страница 2: ...other products are taken into account A decision regarding the combined use of devices from different manufacturers where they do not constitute treatment units is the responsibility of the user The d...

Страница 3: ...e 93 42 EEC EN 62366 2008 EN 60529 2014 ISO 15233 1 2012 Intended use The MultiStim ECO is used to identify and locate peripheral nerves and nerve cords in local and regional anaesthesia for example d...

Страница 4: ...as well as the applicable sub standards Please note that in the worst case all leakage currents and or the patient s aux iliary currents might add up so that inadmissibly high values put the patient...

Страница 5: ...NK accessories Dynamic electric and magnetic interference fields might cause device system interactions They can affect the measurement of the actual stimulation current and in the worst case lead to...

Страница 6: ...n rhythm can be heard i e stimula tion current flows through the patient If the green LED next to the stimulation button is on either the patient circuit is not closed or there is a resistance 12 k in...

Страница 7: ...mulation current intensity 0 5 mA green LED below 0 5 mA is on Volume medium central green LED above the loudspeaker symbol is on Stimulation OFF LED next to stimulation button is off Function keys Bu...

Страница 8: ...e device under any circumstances Send the Multi Stim back to the manufacturer Operation Check and start up Please note Devices showing other reactions must not be put into operation Please contact the...

Страница 9: ...If this is the case you can start using it 5 As soon as you touch the skin surface with the cannula and or puncture the MultiStim ECO starts delivering current pulses You can perceive this visually by...

Страница 10: ...xt to the unequal sign is also on As soon as the circuit is closed and pulses are delivered to the patient the yellow goes out At the same time the green stimulation LED starts blinking and the acoust...

Страница 11: ...ly Only use alkaline manganese batteries e g Varta LR1 4001 LADY N Duracell LR1 LADY N Energizer LR1 LADY N Each time after replacing the batteries check the MultiStim ECO before start up to make sure...

Страница 12: ...CIE Laboratoire Central des Industries Electroniques Aire de la Thur F 68840 Pulversheim Guidelinesandmanufacturer sdeclaration electromagneticinterference acc to EN 60601 1 2 2007 5 2 2 1 table 1 The...

Страница 13: ...le 2 The MultiStim ECO is intended for use in electromagnetic environment as spe cified below Customers or users of the MultiStim ECO must ensure that it is operated in such an environment Immunity te...

Страница 14: ...rt inter ruptions and variations of supply voltage according to IEC 61000 4 11 5 Ur 95 dip of Ur for period 40 Ur 60 dip of Ur for 5 periods 70 Ur 30 dip of Ur for 25 periods 5 Ur 95 dip of Ur for 5 s...

Страница 15: ...3 Veffective value 150 kHz to 80 MHz within ISM bandsa Not appli cable Portable and mobile radio devices should not be used closer to the MultiStim ECO and its cables than the recommended safety dis t...

Страница 16: ...ty that mobile portable communication devices cause interference if they are unintentionally brought into the patient s area For this reason an additional factor of 10 3 is used for calculating the re...

Страница 17: ...g Maximum Output Charge per pulse Q2 0 mA I t Q2 0 mA 2 0 mA 100 s Q2 0 mA 0 2 As Maximum Average Current across a 500 ohm resistive load and at the maxi mum output setting 2 0 mA Iaverage I tw TStim...

Страница 18: ...GmbH Medizintechnologie Karl Hall Strasse 1 78187 Geisingen Germany Key to symbols used in labelling Manufacturer Consult instructions for use Catalogue number Rx only Dispensing with prescription on...

Страница 19: ...19 PAJUNK GmbH Medizintechnologie Karl Hall Strasse 1 78187 Geisingen Germany Phone 49 0 7704 9291 0 Fax 49 0 7704 9291 600 www pajunk com XS190244B_Englisch 2020 01 21 0124...

Отзывы: