Setting temperature
Setting heating time
Setting force
Summary
Sample removal
Set temperature according to mounting material recommendations (see Tables 4
-
1 to 4
-
5 on the
following pages).
Set heating time according to mounting material recommendations (see Tables 4
-
1 to 4
-
5 on the
following pages).
Set force according to mounting material recommendations (see Tables 4
-
1 to 4
-
5 on the following
pages).
1.
Select and install required mold assembly.
2.
Insert sample and add mounting material.
3.
Select temperature, heating time, and force.
4.
Activate mounting operation by pushing start button.
Upon completion, the display will state
“
Cycle Completed
”.
Turn bayonet cover clockwise to open
position and raise ram to remove completed mount.
!
Caution: Bayonet closure and upper ram may still be very hot.
Recommendation:
Clean ram before producing next mount and remove and clean mold assembly
after every 40 hours of operation.
18