General Power 

Requirements

Input power and grounding must be available prior to system installa-
tion. Sufficient AC power must be available for all equipment. A sepa-
rate earth safety ground connection is also required for the whole 
system. 

Please note that an electrical conduit is not an acceptable separate 
earth ground path. However, water pipes or other known good earth 
ground paths may sometimes be used. Consult your local code and 
regulations to determine if water pipes or other earth ground paths are 
acceptable forms of grounding. 

Locate the fuse or circuit-breaker-protected AC outlets and safety 
ground connections as close to the system as possible (in order to 
keep cable runs as short as possible).

System Input 

Power 

Requirements

WARNING! ELECTRICAL HAZARD.

 Grounding circuit continuity is 

vital for the safe operation of equipment. Never operate equipment 
with the grounding conductor disconnected. Wiring must conform to 
country and local electrical codes.

DANGER! ELECTRICAL SHOCK HAZARD

. Severe electrical shock 

can result from operating the instrument without the instrument panels 
in place. Do not remove the instrument panels. High-voltage contacts 
are exposed when instrument panels are removed from the instru-
ment.

DANGER! ELECTRICAL HAZARD. 

Use properly configured and 

approved line cords for the voltage supply in your facility.

DANGER! ELECTRICAL HAZARD. 

Plug the system into a properly 

grounded receptacle with adequate current capacity.

Overvoltage

 

Rating

The Sequel System has an installation (overvoltage) category of III 
(three)

 

and is classified as nonportable equipment.

Electrical 

Connections

The Sequel System must be plugged into a grounded circuit capable of 
delivering at most 30 Amps for a 200-240V power source. 

The maximum power consumption is 3000 Watts. 

DANGER!

ELECTRICAL SHOCK AND LASER HAZARD Installation, mainte-
nance and repair are only allowed for authorized service personnel. Do 
not remove the panels of the Sequel System.

For more information, see the 

Pacific Biosciences Sequel System Site 

Preparation Guide

Содержание Sequel

Страница 1: ......

Страница 2: ...R INDIRECT DAMAGES IN CONNECTION WITH OR ARISING FROM THIS DOCUMENT WHETHER OR NOT FORESEEABLE AND WHETHER OR NOT PACIFIC BIOSCIENCES IS ADVISED OF THE POSSIBILITY OF SUCH DAMAGES Certain notices term...

Страница 3: ...lso be used to alert against unsafe practices WARNING Indicates a potentially hazardous event that if not avoided could result in death or serious injury DANGER Indicates an imminently hazardous event...

Страница 4: ...ent with caution ad only under professional environments The RCM label indicates the product is compliant with the applicable technical standards for Electromagnetic Compatibility EMC The RCM mark is...

Страница 5: ...e presence of an electrical shock hazard and or high leakage current Earth connection is essential before connecting supply Proceed with appropriate caution Environmental Symbols on Instrument Symbol...

Страница 6: ...re to skin Do not operate the instrument with any of the panels removed When the front door of the instrument is open the laser beam is blocked via safety interlock switches If you operate the instrum...

Страница 7: ...proval of the Fire Marshall or other authority having jurisdiction The installation of appropriate fire extinguishers shall be in addition to other fire protection systems and not as a substitute or a...

Страница 8: ...site is appropriately stocked with the needed materials Note that only supported consumables and accessories may be used with the Sequel System In cases of limited user accessibility to the instrumen...

Страница 9: ...mptied by the user The system does not generate any additional waste About Safety Data Sheets Chemical manufacturers supply current Safety Data Sheets SDS with shipments of hazardous chemicals to new...

Страница 10: ...for container storage Handle chemical wastes in a fume hood After emptying a waste container seal it with the cap provided Dispose of the contents of the waste tray and waste bottle in accordance wit...

Страница 11: ...must conform to country and local electrical codes DANGER ELECTRICAL SHOCK HAZARD Severe electrical shock can result from operating the instrument without the instrument panels in place Do not remove...

Страница 12: ...The system must be installed and maintained by a Pacific Biosciences representative All instrument panels must be in place on the instrument while the instrument is operating When all panels are insta...

Страница 13: ...lasers that can damage the eyes Never look directly into the laser beam The scanner is designed to prevent human access to harmful levels of laser light during normal operation user maintenance or dur...

Отзывы: