C.4.  Once the level inside the BBL is balanced and the flowmeter reads 5 LPM, plug-

in the ozone generator to the power outlet and start with the main switch. I high-
pitch sound will tell the ozone-generating cell inside is active and producing 
ozone.  

 
 
D. 

OPERATION. 

 

The system must be set to work automatically. The controlled power outlets for 

the ozone generator and oxygen concentrator must have proper interlock to the 
hydraulic booster pump and other customer controls. Once the suction of ozone stops, 
also the ozone generator and oxygen generator must stop.  

 
 

E.    

MAINTENANCE.  

 

The Ozone Systems are designed to require minimal maintenance.  

THERE IS 

DANGEROUS HIGH VOLTAGE INSIDE!  DO NOT OPEN THE UNITS; 

THERE ARE 

NO SERVICEABLE PARTS INSIDE.

 DO NOT REPLACE FUSES; 

they do not fail 

without good reason. Only trained and qualified personnel are permitted to repair the 
equipment.  The following list of procedures should be performed: 
 
 
DAILY

 
E.1. 

The fans must be running and the airflow unobstructed. If FUSE 1 is lit, call for 
service. 

 

 
E.2.    The moisture indicator in the oxygen line should remain BLUE all time. If 

the color changes, the new color will indicate: 

 

.    

WHITE -  The dew point is above the allowable -40

°

C.  If a power failure took 

place within last five hours, continue to operate as is. The crystals will turn blue 
again in the next five hours. If not, contact your nearest dealer.  

  
 .    

BLACK - The blue crystals are permanently damaged by an exposure to 
ozone.  
Check and restore the room ventilation, replace the indicator. 

 

.    

YELLOW - The blue crystals are permanently damaged by an exposure to   oil 
vapors.  
In this case also the oxygen generator is likely to be damaged and must 
be replaced.  

  
.    

PINK/RED - The blue crystals are permanently damaged by an exposure to 
acid vapors 
in the air. The source of the problem must be identified and 
removed. Examples: A solid fuel burner near by, overcharged battery,  

 
E.3. 

Check the flowmeter for proper airflow. A low flow may indicate a problem with 
the pump, injector, or integrity of the tubing.  

 
E.4. 

Check the BBL to make sure its water level is equal or almost equal on both 
sides.  

 