OUV-2800-HF(T), OUV-4000-HF(T)  &  OUV-6800-HF(T)

Tube Removal & Cleaning Instructions

                                                   DISCONNECT POWER BEFORE SERVICING

                                            DO NOT LOOK AT UV LIGHT

                                                      DO NOT OPERATE WITH end cap OFF!!

1. Remove the 18 Philips (star) screws form the

top lid of the UV machine as shown. 7 on each

side and 4 on the top.

Do not remove the lower screws!

2. Lift the

lid off as

indicated in

the image.

3. Clip the 4

cable ties that

are holding the

tubes in place.

NOTE

 

 

 that if

you plan on

shipping the

unit these ties

MUST be

replaced.

4. To remove the tube lift the end and pull
while wiggling the tube back and forth being

careful not to break it on the steel support /
light guard.

If you have a UV 2700 go to step 5a. 

If you have a UV 4000 go to step 5b.

5a. To remove the

lower level of tubes
on the UV 2700

simply remove the
ties and tubes as in

step 4.

Now go to step 6.

5b. To remove the second and third row of

tubes in the UV 4000 you will need to first
remove the tube support brackets after

removing the tubes. This bracket is
removed by removing 3 screws from each

as indicated in the image. Then remove
the ties and tubes as in step 4.

Now go to step 6.