background image

8+9)

.  With a damp cloth, (not wet) wipe out ozone generator chamber, to remove any debris 

which may be present.

 

5.  Dry the plates completely by putting them in an oven for 15 minutes @ 350 deg F.  (Do not 

put them in the microwave!) 

6. Once plates are inspected and dried, they may be reassembled.  Place one mica sheet flat in 

the palm of your hand 

(Figure 10)

, then set the inner screen in the exact center of mica sheet.  

Then  place  the  other  sheet  of  mica  on  top forming  a  sandwich 

(Figure  11)

.   With  thumb  and 

index finger grasp mica sandwich approx. half way 

(be careful inner screen remains centered 

at all times) 

and gently slide assembly back into outer screen 

(Figure 12).

 

 

 

 

 

 

 

 

 

 

 

REPLACING SERVICED PLATE IN MACHINE

 

1.  Install plates in reverse order of removal being careful inner stainless steel screen remains 

centered at all times.   

2.  Once plates have been reinstalled, ensure that the high voltage lead is clamped tightly in the 

holding clip (Figure 4). 

3.  Clean intake filters if dirty, then plug unit in and test for proper operation. 

4.  If the plates are slightly damp, ozone production will be significantly diminished until the inner 

fan dries the plates out.  Let the unit run for 5 minutes and then verify the ozone output by smell. 

 

TROUBLESHOOTING:

 

If  unit  makes  a  snapping  or  popping  sound,  ensure  that  the  plates  are  centered.    If  it  still  makes  a 

snapping  sound,  turn  down  the  variable  control  until  the  snapping  sound  stops.    Snapping  is  more 

common  in  very  dry  climates  since  an  electrical  charge  can  travel  further  in  dry  air.    Ensure  upon 

reinstallation  that  both  inner  and  outer  stainless  steel  are  centered  as  close  as  possible,  and  that  all 

components of generator are clean, dry and damage free. 

 

1.  Fan works but no ozone:   

A: Is the filter plugged? 

B: Are the plates clean and dry?  (refer to maintenance section) if unit does not work after cleaning it will 

probably need to be sent in for service work. 

C: A hole has formed in the corona plate causing a direct short. – Contact Ozone Solutions at 888-892-

0303 if this is suspected. 

2.  Unit snaps or sparks: 

A: Are the plates clean and dry?  (refer to maintenance section) 

B: Is the center screen in the plate centered?  (If screen is off center place back in center of plat, refer to 

maintenance instructions for plate removal and replacement). 

 

 

Figure 10 

 

Figure 11 

 

Figure 12 