background image

OMZ-1000-HF (HFT) 

Plate Removal Instructions

1. The new hose connector/end cap is 

much stronger and has an ozone 

resistant silicone seal. The cap is 

removed the same way as the original

model, by removing 2 screws. 

2.  The generator plate rack is secured

by two screws that are located inside 

the unit behind the end cap instead of 

on the bottom as indicated in the 

original manual. 

3. The plate 

rack is 

removed by 

lifting it up 

and pulling it 

out. 

4. The plates are held in 

place by a bar that is held 

in place by a wing nut, the 

HV contact bar is also held 

down by 2 wing nuts, 

remove these three wing 

nuts.

5. Once the bar and 

contact bracket are 

removed, remove the 

plates by lifting them 

straight up as shown. 

6. Next refer to the plate cleaning instructions. 

PLEASE ENSURE THAT ALL PARTS ARE COMPLETELY DRY PRIOR TO RE ASSEMBLY! 
7. Reinstall plates by 

first using a screw 

driver to separate the 

tight ring springs as 

shown. Then slip the 

plate in between the 

springs, making sure 

that the spring does 

not go up in the 

middle of the plate as 

shown. 

8. Re install the plate hold down bar as shown in the step 4 picture. Re install the HV clamp making sure that all the 

HV wires from the plates are clamped tight. 
9. Please 

note that 

when re 

installing 

the plate 

rack, the 

front of it 

must fit under the screws 

indicated in the diagram. 

10. Replace 

the plate 

rack, reinstall 

the screws 

that hold the 

rack in place 

and reinstall 

the end cap 

and screws. 

11. To access the filter remove the end 

cap at the air input end of the ozone 

generator in the same way the output cap 

was removed in step one. The filter may be 

washed with soapy water or replaced with a

new filter available thru your supplier. If 

another brand of filter is used please ensure

that it has good air flow or else the ozone 

generator could overheat due to lack of 

cooling. 

Содержание OMZ-1000-HF

Страница 1: ...s running The room should not be reentered until all ozone has been depleted unless proper breathing respirators are used MAINTENANCE FREQUENCY Under heavy duty use or severely polluted areas The ozone generator should be inspected and cleaned if necessary every 2 to 3 weeks for fine dust or oily residue collecting on generator surfaces or plates Light duty use requires cleaning every 2 to 6 month...

Страница 2: ... inline with the Hi Low switch This timer is designed to allow the user to turn on the ozone generator and have it turn off after a given amount of time or run it continuously the word HOLD means continuous To use the timer first make sure the switch is in the Hi or Low position a To turn the timer on to constantly run simply press and hold the button for a couple seconds the top green LED will li...

Страница 3: ... RE ASSEMBLY 7 Reinstall plates by first using a screw driver to separate the tight ring springs as shown Then slip the plate in between the springs making sure that the spring does not go up in the middle of the plate as shown 8 Re install the plate hold down bar as shown in the step 4 picture Re install the HV clamp making sure that all the HV wires from the plates are clamped tight 9 Please not...

Страница 4: ...d screens If using an oven set the temperature at about 200 F and place the mica and screens on the middle rack for approximately 1 2 hours If there is no rush or you do not have access to an oven place the mica and screens in a very warm dry place such as on an operating base board heater to dry overnight or if it must air dry leave it dry for several days with good air flow 5 Once all the Mica a...

Отзывы: