Caution Beryllium
The detector has a thin beryllium window. Beryllium is a toxic compound, however, the beryllium
window poses no health hazard when it is intact.
Caution; Toxic Material: Do not puncture, break or damage the beryllium window in any
way. This can produce airborne particles. Prolonged inhalation of beryllium can cause cancer.
Do not allow the detector to come into contact with moisture, or condensation from high humidity.
This can corrode the beryllium window, in particular if chlorine, sulphates, copper or iron is also present.
Contact the local Oxford Instruments representative for advice about beryllium, or if the beryllium
window is pierced, broken, damaged or corroded.
Caution Lanyard Use
The X-MET8000 series includes a lanyard. This is only applicable for use at ground level.
Caution: Do not use the X-MET8000 series lanyard as a safety lanyard for work at height
because of the risk of a fall. This can result in serious personal injury.
The anchor for the lanyard on the underside of the X-MET8000 series is applicable for use with a tool
safety lanyard for work at height.
X-MET8000 Series Safety Features
The X-MET8000 series includes these 10 key safety features to protect the operator.
Power Button And Indicator
Safety Feature: Press and hold the Power button for 5 seconds to switch the X-MET8000
series on or off.
Password Protection
Safety Feature: An operator must have the correct password to use the X-MET8000 series.
Safety Feature: The supervisor can change the passwords.
Proximity Sensors
Safety Feature: The sample must cover the proximity window before the X-MET8000 series
can generate an X-ray beam.
Safety Feature: When the sample covers the proximity window, the proximity indicators
change to orange.
Safety Feature: The X-MET8000 series switches the X-ray beam off if there is no return signal
from the sample.
9
X-MET8000 Series Operator’s Manual
8
Содержание X-MET8000 Series
Страница 1: ...X MET8000 Series Operator s Manual...
Страница 58: ...2 Click New Report The Reports screen has a new report 58 X MET8000 Series Operator s Manual 8...
Страница 59: ...3 Click Load Results The Measurements dialog box appears 59 X MET8000 Series Operator s Manual 8...
Страница 96: ...4 Tap Disconnect to end the VNC session The VNC session ends 96 X MET8000 Series Operator s Manual 8...
Страница 120: ...Taiwan NCC Notice 120 X MET8000 Series Operator s Manual 8...