6.
Measure the distance from the sample when the proximity indicators switch off.
The distance should be no more than 15 millimetres. If the X-MET8000 series does not achieve this
test, it must be returned to the local Oxford Instruments representative for service.
Note: the proximity sensor may be disabled by a Supervisor for operation of the analyser with a bench
top stand only (as per IEC 62495). If unsure, please refer to local radiation safety regulations.
Recommended Weekly Maintenance
Follow these procedures on a weekly basis.
Check The Alloy CRM Sample
The supervisor should make a measurement of the Check sample(s).
If there is a significant change from the Alloy CRM reference measurement, contact the local Oxford
Instruments representative for assistance.
The X-MET8000 series can include additional check samples that are application specific. The supervisor
can also use these to check the X-MET8000 series against a representative sample.
Battery Maintenance
The Power Supply can charge the battery while supplying power to the X-MET8000 series. The
X-MET8000 series display has a battery icon showing Remaining Charge and Charge Status.
Remaining charge is indicated by a battery symbol filled to the calculated charge of the battery.
The Charge Status indicator on the X-MET8000 series display indicates when the battery is charging
by an animation in the battery icon.
When the battery is fully charged the charging animation stops and the charge indicator on the
X-MET8000 series displays a fully charged battery symbol.
If the X-MET8000 series is connected to the Power Supply while turned off, the power button light on
the X-MET8000 series pulsates to indicate that the X-MET8000 series is charging.
The time required to fully charge a battery in the X-MET8000 series is up to 8 hours with the X-MET8000
series switched off.
The time required to fully charge a battery using the Battery Charger is up to 6 hours.
The X-MET8000 series can be operated from the Power Supply without a battery. When connected to
the Power Supply the battery can be removed while the X-MET8000 series is powered up.
If the X-MET8000 series is operated from the Power Supply without a battery, the battery icon is
displayed as empty.
Make sure that the Power Supply has the correct mains adapter, or a mains extension lead.
Always keep the battery charged. Storing the battery in a discharged state can damage the battery
and if left in a discharged state for longer time the cells will enter a deep discharge state where the
internal protection circuit prevents charging for safety reasons.
111
X-MET8000 Series Operator’s Manual
8
Содержание X-MET8000 Series
Страница 1: ...X MET8000 Series Operator s Manual...
Страница 58: ...2 Click New Report The Reports screen has a new report 58 X MET8000 Series Operator s Manual 8...
Страница 59: ...3 Click Load Results The Measurements dialog box appears 59 X MET8000 Series Operator s Manual 8...
Страница 96: ...4 Tap Disconnect to end the VNC session The VNC session ends 96 X MET8000 Series Operator s Manual 8...
Страница 120: ...Taiwan NCC Notice 120 X MET8000 Series Operator s Manual 8...