Version 1.4 rev 23 Apr 2019
20
IKon-L
3.6 c
onnecting
tHe
i
k
on
-l
to
tHe
p
oWeR
s
Upply
1. Connect the power cable between the camera power input and the PS-40, ensuring it is threaded correctly and
tigthened fully.
2. Ensure the PS-40 is switched off and then connect the power cable of the PS-40 to the PS-29.
3. Connect the PS-29 to the mains supply.
4. To connect a water/liquid cooling system proceed to
Section 3.7.
For air cooling applications proceed to
Section 3.10
to install the software.
Notes
•
Ensure that the power connectors for the iKon-L camera are inserted securely. The connectors are
keyed to aid correct orientation.
•
The
symbols on the power cable, the PS-40 and the camera must be aligned correctly:
•
If a replacement 16-pin cable (Andor part No.: ASE-01957) is required please contact your nearest Andor
representative.
•
The electrical mains lead should be certified for use in your country and in applicable countries the plug
must be fitted with a 240V 5A fuse.
PS-29
PS-40
iKon-L
Connect to the
Mains supply
PS-40
PS-29
16-pin connection
cable A
SE-01957
Connect the 4-pin
output cable of
the PS-29 to the
PS-40 input.
Note the alignment arrows
during insertion!
Note the alignment arrows
during insertion!