Clara
23
Version 2.0 rev 24 Mar 2015
OPERATION
4.6.4 a
CCumulate
C
yCle
t
ime
& n
o
.
of
a
CCumulations
If you have selected Accumulate or Kinetic as the acquisition mode, with Internal triggering, you can also select the
Accumulation Cycle Time and No. of Accumulations.
The Accumulation Cycle Time is the period in seconds between each of a number of scans, whose data are to be added
together in computer memory to form an Accumulated Scan.
The Number of Accumulations indicates the number of scans you want to add together.
4.6.5 k
inetiC
s
eRies
l
engtH
& k
inetiC
C
yCle
time
When Kinetic is selected as the acquisition mode, with Internal triggering you can also select the Kinetic Series Length
and Kinetic Cycle Length (secs).
•
Kinetic Series Length
: the number of scans you require in your series
•
Kinetic Cycle Length
: the interval (in seconds) at which each scan (or accumulated scan) in your series begins