5.Panel Operation
54
(1)
Press the front panel
Utility
function key, press the
System
softkey.
(2)
Press the
Beeper
softkey to toggle between
On
or
Off
.
Clock Source
The generator provides an internal 10MHz clock source and can receive the external
clock source from the
10MHz/In/Out/Counter
connector at the rear panel. It can
also output an internal clock source from the
10MHz/In/Out/Counter
connector for
other devices.
Note:
The amplitude range of the
10MHz In/Out/Counter
input signal is 1 Vpp to 3.3 Vpp.
(1)
Press the front panel
Utility
function key, press the
System
softkey.
(2)
Press the
CLK Ref
softkey to toggle between
Internal
and
External
.
Note:
The clock source defaults to the internal clock source. When an external clock source
is required, this function is switched to external. At this time, the clock output is
turned off forcibly.
Clock Output
(1)
Press the front panel
Utility
function key, press the
System
softkey.
(2)
Press the
CLK Output
softkey to toggle between
On
and
Off
.
Note:
When the clock source output is turned on, the clock source must be switched to
internal, and the counter function stops after the clock output is turned on.
Firmware Update
(1)
Press the front panel
Utility
function key, press the
System
softkey.
(2)
Plug the USB storage device into the USB connector on the front panel of the
instrument.
Note
: If the USB storage device is not inserted, the
Update firmware
menu is
disabled.
(3)
Press the
Update firmware
softkey, turn the
knob
to select
USBDEVICE
, press
the
Enter
softkey to enter the USB storage device to browse files.
(4)
Turn the
knob
to select the downloaded firmware file, and press the
Execute
softkey to perform firmware upgrade.
Note
: