4.Quick Start
10
The supplied accessories have been already described in the "Appendix A: Enclosure"
of this Manual. You can check whether there is any loss of accessories with reference
to this description. If it is found that there is any accessory lost or damaged, please get
in touch with the distributor of Lilliput responsible for this service or the Lilliput's
local offices.
3. Check the Complete Instrument
If it is found that there is damage to the appearance of the instrument, or the
instrument cannot work normally, or fails in the performance test, please get in touch
with the Lilliput's distributor responsible for this business or the Lilliput's local offices.
If there is damage to the instrument caused by the transportation, please keep the
package. With the transportation department or the Lilliput's distributor responsible
for this business informed about it, a repairing or replacement of the instrument will
be arranged by the Lilliput.
4.3
Power-On Check
4.3.1
AC Power Input Setting
ODP3031 adopt 110V/220V AC power source. Users should regulate the voltage scale of
the
Power Switch
according to the standards in their own country (see
Figure 4-2
) at the
rear panel and use an appropriate fuse.
Voltage
ODP3032
AC110V
125 V
,
F5 A
AC220V
250 V
,
F3 A
To change the input voltage scale of the instrument, do the following steps:
(1)
Turn off the power button at the front panel and remove the power cord.
(2)
Check if the fuse installed before leaving factory (250 V, F3 A) can match with the
selected voltage scale; if not, pry the cover open using a straight screwdriver (see
⑤
in
Figure 4-2
), change the fuse.
(3)
Regulate the
Power Switch
to the right voltage scale.
4.3.2
Power On
(1)
Connect the instrument to the AC supply using the supplied power cord.
Warning:
To avoid electric shock, the instrument must be grounded properly.
(2)
Press down the
power button
at the front panel, the orange and blue key is lighted;
the screen shows the boot screen.
(3)
Press any key to enter.