User’s Manual of
OWON
Color Mixed Signal Digital Storage Oscilloscope
General
Deep memory
512 ms
8 s
General
Deep memory
5120 s
80000 s
200kHz Low
memory
General
Deep memory
80 ms
1.28 s
20 s
20Hz Low
memory
General
Deep memory
800 s
12800 s
200000 s
100kHz Low
memory
General
Deep memory
160 ms
2.56 s
40 s
Follow up steps to set the measurement for data bus of 33M clock:
Sample rate: 100M Storage depth: 4M
Filter: none
Setting steps as below:
1. Press “
E(ACQUIRE)
” and sample menu appears.
2. Press “
F1
”repeatly or turn “
CH1 Volts/div
” knob till sample rate set as “100M”
3. Press “
F2
” till storage depth display as “
Deep Memory
”.
4. Press “
F3
” till digital filter display as “
None
”.
Then sample system setting finished (refer to fig 61).
Fig 61
: Sampling setting
How to set trigger system
Trigger system including “
Trigger level
” adjust knob, “
Trigger menu
”, “
SET 50%
”,
“
Set to Zero
”. Trig level adjust knob is to adjust trigger position. Since the digital storage
depth is fixed so the proportion of storage data between pre-trigger points and re-trigger
ones depend on trigger position. Trigger mode including: Edge trigger, bus trigger, pattern
73