User’s Manual of
OWON
Color Mixed Signal Digital Storage Oscilloscope
Pattern trigger menu as below:
Function
Setting
Instruction
Channel
choice
CH00
~
CH0F
x0100000
00000000
Select the channel to set signal pattern.
16 channel status indicator.
X
:
don
’
t care
0
:
low
1
:
high
CH00
~
CH0F Don
’
t Care
Low
High
No trigger occur.
Trigger occur when signal on the low level
Trigger occur when signal on the high level.
To observe the 16 channel data need to know how data changed before and after
when "BIT0-BIT3" STATE is “0111”.
A. Press
“
Trigger menu
”
and menu appears.
B. Press
“
F1
”
till trigger mode display as
“
Pattern
”
C. Press “
F2
” or turn “
CH1 Volts/div
” knob till “
CH SEL
” display as “
CH00
”
D. Press “F3” till F3 window display as “CH00/High” and “CH00” in F2 window
display as “
1
”.
E
、
Repeat operation of step D and set CH01, CH02 to
“
High
”
and CH03 as
“
Low
”
,CH04~CH0F set as
“
Don
’
t care
”
.
Pattern trigger setting finished (refer to fig. 67) and ready for data acquisition.
Fig 67
:
Pattern trigger
4
.
Sequential queue trigger
: make BUS as trigger source and continuous setting
data in BUS as trigger condition to generate trigger and also can set 8 data at the
same time. See Fig 68
71