How to set sampling system
Sampling system is to set sample rate, storage depth and filter. Different sampling
setting will result in different measure results. In the same storage depth, the
higher sample rate set, the shorter the continuance time for signal will be. If the
sample rate set too low, narrower pulse signal may get lost. So the sample rate
and storage depth should be set correctly according to the actual measuring signal.
Press
“
E (ACQU)
”
and display menu as the figure 60:
Figure 60
:
Sampling menu
Sample menu function list:
Function Setting Instruction
Sample
rate
In sequence of
1-2-5 from 20Hz
to 100MHz
Storage
depth
Low memory
General
Deep memory
Storage depth of 16K
Storage depth of 256K
Storage depth of 4M
Digital
Filter
None
1
2
Filter closed
Filter one pulse of sample width
Filter two pulse of sample width
Listing of corresponding continuance time to different sample rate and storage
depth: