.Menu Interpretation
43
the frequency of current marker.
In Delta marker mode, the center frequency step will be set to
the frequency at which the Delta Marker is located.
The function is not available in zero span mode.
5.6.3.3 [Mkr->Start]
Sets the start frequency of the analyzer to the frequency of the
current marker.
In Normal marker mode, the start frequency will be set to the
frequency of the current marker.
In Delta marker mode, the start frequency will be set to the
frequency at which the Delta Marker is located.
The function is not available in zero span mode.
5.6.3.4 [Mkr->Stop]
Sets the stop frequency of the analyzer to the frequency of the
current marker.
In Normal marker mode, the stop frequency will be set to the
frequency of the current marker.
In Delta marker mode, the stop frequency will be set to the
frequency at which the Delta Marker is located.
The function is not available in zero span mode.
5.6.3.5 [Mkr->Ref Level]
Sets the reference level of the analyzer to the amplitude of the
current marker.
In Normal marker mode, the reference level will be set to the
amplitude of the current marker.
In Delta marker mode, the reference level will be set to the
amplitude at which the Delta Marker is located.
5.6.3.6
[MkrΔ
->Span]
Sets the span of the analyzer to the frequency difference between
the two markers in Delta marker mode.