Optical Power Meter:
ZOOM 2
Fiber Optic Light Source: Dual OWL Series
PART 1 ACCESSORIES REQUIRED FOR THIS PROCEDURE
–
Please note: accessories will vary based on factors such as core size (62.5/125 vs. 50/125), test equipment connector type, and
patch panel connector type. The diagram below should help visualize when selecting the correct accessories to use.
A helpful video regarding choosing patch cables is available on OWL’s website:
https://owl-inc.com/videos/htm/concepts/c1011.htm
ZOOM 2
Dual OWL Series
(850nm only or 850/1300)
(SC or ST)
KF7X
DETECTOR
FIGURE 3
–
Required Accessories
50/125 µm
62.5/125 µm
Diameter
Wraps
0.9 in.
0.7 in.
5
5
DETECTOR NOTE:
For testing LC patch panels, install the 1.25mm adapter cap
For testing SC, ST, or FC patch panels, install the 2.5mm
adapter cap
NOTE: each light source port requires its own reference cable
REFERENCE CABLE (
Œ
) & MANDREL (
)
– OR –
EF
EF MODE CONTROLLER CABLE (
)
62.5/125 – 0.7” mandrel
50/125 – 0.9” mandrel
MATING SLEEVE (
Ž
)
TEST CABLE (
)
PART 2 SET REFERENCE
–
The ZOOM 2 display should change to show readings in ‘dB’, and the power reading should be approximately
0.00 dB (Figure 7). This indicates that the meter and light source have been successfully “zeroed” at 850nm.
NOTE: If also testing at 1300nm, connect the 1300nm reference cable to the ZOOM 2, set both units to
1300nm, then repeat this step.
Press
to power on the Dual OWL. The 850nm indicator LED should be lit by default.
Note: if Encircled Flux compliance is required, use Encircled Flux Mode Controller Cables instead (
).
Press
to power on the ZOOM 2.
Press
to set the ZOOM 2 to the 850nm wavelength.
Connect the ZOOM 2 and Dual OWL together using the mandrel-wrapped patch cable (
Œ
&
) as shown in Figure 5.
The ZOOM 2 should read approximately -20.00 dBm for 62.5/125 OM1 fibers, or around -23.00 dBm for
50/125 OM2/3/4 fibers (Figure 6).
NOTE: each light source port requires its own reference cable
Press and hold
to set the reference for 850nm.
UNIT S
Z E R O
FIGURE 5
– Connect meter and source with reference cable
dBm
20.00
-
850
nm
FIGURE 6
dB
0.00
850
nm
FIGURE 7
REFERENCE CABLE (
Œ
) & MANDREL (
)
Œ