OPERATION/MAINTENANCE
UTILITIES MENU
From any test results screen on the MASTER, press the MENU button to access the MAIN MENU.
From the MAIN MENU, select SETUP MENU, then select UTILITIES MENU.
SET SYSTEM CLOCK
Set real-time clock
SET APD BIAS
Set avalanche photodiode bias voltage
FORMAT DATA FLASH
Format data storage
FACTORY RESET
Reset to factory defaults
MANUFACTURER SETUP
Not used
PROGRAM CPLD
Not used
VIEW SLIDES
View slide graphics
UTILITIES MENU
SET SYSTEM CLOCK
SET APD BIAS
FORMAT DATA FLASH
FACTORY RESET
MANUFACTURER SETUP
PROGRAM CPLD
VIEW SLIDES
SELECT
EXIT
BACK
Set real time clock
UTILITIES MENU
ð
SET SYSTEM CLOCK
SET SYSTEM CLOCK
– set time and date
DATE
Format: MM/DD/YY (MM=month/DD=day/YY=year)
TIME
Format: HH:MM:SS (HH=hour/MM=minute/SS=second) : AM/PM
NEXT / PREVIOUS field
DECREASE / INCREASE currently selected value
Save settings and return to UTILITIES MENU
Exit without saving settings and return to SETUP MENU
SET
EXIT
Enter Date (MM/DD/YY)
Enter Time (HH:MM:SS)
SET
EXIT
10:12:19:AM
4- 4-18
UTILITIES MENU
ð
SET APD BIAS
SET APD BIAS
– set avalanche diode bias voltage
Sets a baseline voltage for the OTDR avalanche photodiode.
Under normal operation the voltage percentage should count up to 70-80%.
If it does not, contact OWL technical support for more information.
SETTING APD BIAS VOLTAGE
BIAS VOLTAGE % : 72