TEST PROCEDURES
LINK WIZARD
After a few seconds, the SOURCE REFERENCE POWER will be shown. Press
to continue.
DONE
Once the WaveSource Pro is set to AUTO mode, press
on the Fiber OWL 7+ to
set the SOURCE REFERENCE POWER.
TEST
Optical Source
WaveSource Pro
AUTO
PORT
MOD
CW
Optical Source
WaveSource Pro
AUTO
PORT
MOD
CW
Wavelength
Set source to AUTO or 1310nm
You will need to set a source
power reference for each
wavelength that you are testing.
Press TEST to measure.
Press DONE after testing all
wavelengths.
Testing...
Power
Typical
TEST
WAVE
BACK
DONE
SOURCE REFERENCE POWER
HELP
Wavelength
1310nm
1550nm
Set source to AUTO or 1310nm
You will need to set a source
power reference for each
wavelength that you are testing.
Press TEST to measure.
Press DONE after testing all
wavelengths.
TEST
WAVE
BACK
DONE
SOURCE REFERENCE POWER
HELP
Power
-10.20dBm
-10.04dBm
Typical
-10.00dBm
-10.00dBm
Note about Typical power values:
If the reference power levels are over 1 dB below
the typical value for that fiber type, the patch
cable either needs to be cleaned or replaced.
STEP 4: SETTING THE REFERENCE, cont.