background image

Appendices

UNIT 3

3-1

Maintenance and Calibration Procedures

Specifications

Contact Information

Repair.  

Repair of this unit by unauthorized personnel is prohibited, and will void any warranty associated with 

the unit.

Battery Replacement. 

 The battery compartment is covered by a sliding plate on the back of the unit.  One 9v 

battery is required for operation.

Cleaning.  

For accurate readings, the optical connectors on the BOLT and the connectors on the patch cords 

should be cleaned prior to attaching them to one another.  Minimize dust and dirt buildup by replacing the dust 
caps after each use.

Re-calibration.

   It is recommended to re-calibrate the BOLT every 12 months to maintain the light source 

stability at -10 dBm.

Warranty.  

The BOLT comes standard with a one-year factory warranty, which covers manufacturer defects 

and workmanship only.

Address:
Optical Wavelength Laboratories
N9623 Hwy 12
Whitewater, WI 53190

Phone:
(262) 473-0643

Internet:
OWL-INC.COM

Launch Method:
Output Power (62.5 µm core):
Resolution:
Accuracy:
Measurement Range:
Battery Life:
Operating Temperature:
Storage Temperature:
Low Battery Indicator:
Connector Style:
Width:
Height:
Depth:
Weight (with battery):

FP Laser

0.1 milliwatts

up to 0.001 kilometers

± 2.5 meters

25 kilometers

10 hours

0 to 55° C
0 to 75° C

Yes

ST

2.75 inches
4.94 inches
1.28 inches

154 grams

Содержание Bolt

Страница 1: ...S Hwy 12 Whitewater WI 53190 Phone 262 473 0643 Internet OWL INC COM Date July 26 2004 Version 1 1 BOLT Operations Guide Beaming Optical Length Tester Optical Wavelength Laboratories The s W e i is r e o t C a h r o o i c b e a L i n h t O g p n t e i l c e a v l a F i b W e l r aci T t e p s t O OWL ...

Страница 2: ...ed Accessories 1 2 Applications 1 2 Unit 2 Operation Fiber Length Measurement 2 1 Fiber Identification 2 1 Single mode Light Source 2 2 Unit 3 Appendices Maintenance and Calibration Procedures 3 1 Specifications 3 1 Contact Information 3 1 Conversion Factors 3 2 Table of Contents ...

Страница 3: ...so includes a pulse mode feature for fiber location when used with a fiber identifier as well as a continuous wave mode to be used as a stable 10 dBm single mode light source Description 1 1 Introduction UNIT 1 1310NM FP LASER TRANSMITTER this ST connector port houses a laser diode that emits invisible light into an optical fiber CONTINUOUS WAVE TEST SELECTOR SWITCH this 3 way switch selects betwe...

Страница 4: ...required to loop back the two fibers being used for the test The connector styles on the patch cords must match the type on the BOLT and the type of the bulkheads in the system under test Applications Fiber Length Measurement The BOLT uses a round robin method of length measurement This method requires two fibers from the same fiber cable to be looped back at one end with a patch cord Light is inj...

Страница 5: ... right the link is 1 250 kilometers NOTE If the green indicator LED does not light up this means that the battery power insufficient for a valid test The battery must be replaced before continuing Fiber Identification 1 Connect the BOLT to the fiber you are testing via a patch cord as shown at right 2 Power on the BOLT into fiber identification mode by flipping the switch to TEST This mode emits a...

Страница 6: ...ter such as the Fiber OWL to the other end of thefiber you are testing via a patch cord as shown at right 3 Power on the BOLT to continuous wave light source mode by flipping the switch to CW Allow the BOLT to stabilize for 2 minutes before taking measurements with the power meter 2 2 Patch Cord Fiber Link Single mode Light Source To power meter ...

Страница 7: ...acing the dust caps after each use Re calibration It is recommended to re calibrate the BOLT every 12 months to maintain the light source stability at 10 dBm Warranty The BOLT comes standard with a one year factory warranty which covers manufacturer defects and workmanship only Address Optical Wavelength Laboratories N9623 Hwy 12 Whitewater WI 53190 Phone 262 473 0643 Internet OWL INC COM Launch M...

Страница 8: ...1 250 x 1000 meters 1250 meters kilometers to miles 1 250 x 0 6214 miles 0 777 miles kilometers to feet 1 250 x 3281 feet 4101 25 feet Appendices UNIT 3 3 2 Conversion Factors The BOLT displays fiber length in kilometers At times it may be necessary to convert the fiber length to a different unit of measurement The table below provides conversion factors for various units of measurement To calcula...

Отзывы: