ANNEX 5
A5. ELECTROMAGNETIC COMPATIBILITY
Electromagnetic compatibility (EMC) is assessed with reference to the following standards:
IEC/EN 60601-1-2: 2014 (4 edition)
EMISSION
• CEI EN 55011: 2013
• CEI EN 61000-3-2: 2015
• CEI EN 61000-3-3: 2014
IMMUNITY
• CEI EN 61000-4-2: 2011
• CEI EN 61000-4-3: 2007 + A1: 2008
• CEI EN 61000-4-4: 2013
• CEI EN 61000-4-5: 2007
• CEI EN 61000-4-6: 2011
• CEI EN 61000-4-8: 2013
• CEI EN 61000-4-11:2006
Guidance and manufacturer’s declaration – electromagnetic emissions
Owandy-RX DC
is intended to be used in the electromagnetic environment specifi ed below. The customer or the
operator of
Owandy-RX DC
must ensure that the device is used in this type of environment.
Emission test
Conformity
Electromagnetic
environment guidance
RF emissions
CISPR 11
Group 1
Owandy-RX DC uses RF energy only for internal
operation. RF emissions are extremely and
attenuated are not likely to generate interference
with electronic equipment in the vicinity.
RF emissions
CISPR 11
Class B
Owandy-RX DC is suitable for use in all establishments,
including domestic establishments and those
directly connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
CEI EN 61000-3-2
Class A
Voltage fl uctuations/fl icker
emissions
CEI EN 61000-3-3
Complies
NOTE:
Ensure that the device is not stack and location close to other EQUIPMENT please refer to the «
Recommended
separation distances between portable and mobile RF communication equipment and Owandy-RX DC medical
device»