OWANDY RX - HUMAN VERSION
OWANDY RX - HUMAN VERSION
OWANDY RX - HUMAN VERSION
Page 22 of 61
OPERATOR’S MANUAL • Owandy-RX DC • 02/2020 • NXDCEN160A
Operator’s manual – Owandy-RX DC
STEP 2 : check the selected radiographic distance CONE
The LED of the cone length (source-skin distance = SSD) in use should be turned on
LED
8” ON
indicates that the selected tubehead is equipped with 8” = 20cm (SSD) cone
LED
12” ON
indicates that the selected tubehead is equipped with 12” = 31cm (SSD) cone
to change the selection call the “Assistance Service”
PLEASE NOTE
After the modifi cation, default exposure values will be automatically changed.
STEP 3 : check the selected type of tubehead
The LED of the type of selected tubehead should be turned on
LED
AC ON
indicates that the selected tubehead works in alternate current technology
LED
DC ON
indicates that the selected tubehead works in direct current technology
It is not possible to change the selection: an dc device has to be used only in combination with an dc tubehead
STEP 4 : check the selected radiographic voltage
The LED of the radiographic voltage should be turned on
LED
60kV ON
indicates that the radiographic system is set with the high contrast radiodiagnostic technology
LED
70kV ON
indicates that the radiographic system is set with the low contrast radiodiagnostic technology
to change the selection press again the button