background image

E-73-2-E

3

(1)  Mechanical  vibration,  shock  and  corrosive 

gases least exist.

(2) Air is dry and temperature at room temperature

and stable.
NOTE: Although the manufacturer guarantees 
stated  performance  at  ambient  temperatures
from -10 to +50℃ ,  it is recommended that the 
instrument  be  placed  in  service in a room 
temperature environment.

(3)  Potential  sources  of  inductive  interference, 

such as electromagnetic contactors, are locat-
ed sufficiently away.

(4)  A  lightening  arrestor  is  provided  if  incoming 

signals  are  subject  to  potential  influence  of 
lightening.

1. GENERAL

This distributor accepts a train of incoming pulses from the flowmeter of most kinds, or other sig-
nal generating sources, shapes its wave form and, with desired pulse width, furnishes three sets 
of mutually isolated outputs, isolated also from the input.

2. FEATURES

(1) Compactly built plug-in design. Easy to use and maintain.
(2) Provides a power supply to the pulse generator and preamplifier (converter) of a variety of
     flowmeters.
(3) A free type power supply accepts 85-265VAC power source (or 20 to 30V DC in the DC version).

3. INSTALLATION

Installation Location

Select an installation location where

(5) In an application where two or more distributors are to be installed side by side, a sufficient 

working space is secured behind the instrument to facilitate wiring and maintenance (see Fig. 3.1).

(6) For use in an environment where noise level is high, we recommend to use commercially 

available noise filters, isolation transformers, or similar devices.

Dimensions in millimeters

2-M4 Mounting Hole

Fig. 3.1 Socket Mounting Centers

min.140

40 ± 0.2

40 ± 0.2 min.

20

Fig. 3.2 Distributor Outline Dimensions

Fig. 3.3 Socket Outline Dimensions

Lock Lever

Terminal Screw: 11-M3.5 × 7

Содержание SU1308

Страница 1: ...its design performance throughout the life of this distributor this manual offers the operator the necessary installation operation and maintenance information Be well familiar with these instruction...

Страница 2: ...manual are the signal words NOTE CAUTION and WARNING as described in the examples below NOTE Notes are separated from the general text to bring the user s attention to important information CAUTION C...

Страница 3: ...lly isolated outputs isolated also from the input 2 FEATURES 1 Compactly built plug in design Easy to use and maintain 2 Provides a power supply to the pulse generator and preamplifier converter of a...

Страница 4: ...ed Transmission length also depends on the specification of that instrument For other signal cables electrostati c a l l y shielded polyethylene insulated vinyl sheathed control cables CEVS 1 25 to 2...

Страница 5: ...n 3 Pulse generator valid on the pulse generator select board 1 Input wiring is incorrect 2 Pulse generator is at fault 3 Pulse generator select board is incorrect 4 Power supply board or pulse g e n...

Страница 6: ...OPEN 0 3 to 12ms CLOSE 5 0 to 250ms TP1and 0VB If output mode is Duty sync with Input 1 1 it adjustment is correct at 1 ms 8 INDIVIDUAL JUMPER SETTING Input Setting Establish input pulse mode Establis...

Страница 7: ...ance during ON 16 max Leakage current during OFF 1 A max Output pulse width One shot 1ms 50ms Upper cutoff in width of pulses sync with input 10ms 100ms Duty sync with input 1 1 Corresponds to output...

Страница 8: ...pen collector pulse Other than above Output Pulse Width Pulse width 1ms Pulse width 50ms Sync with input Upper cutoff in width of pulses 10ms Sync with input Upper cutoff in width of pulses 100ms Sync...

Отзывы: