D-555-3-E
9
8.2 Parameter Setup Procedure
8.2.1 Procedure to modify a parameter
Given below is the parameter setup procedure:
(1) In "Measure Mode (normal mode)", turn MODE switch ON for 5 seconds to go into "Review Mode".
(2) Using MODE and RESET switches, show the parameter you want to modify.
(3) Turn MODE switch ON for 2 seconds to go into "Parameter Setup Mode".
(4) Using MODE and RESET switches, set up a new parameter. (See Paragraph 8.2.2 for the procedure.)
(5) Following the parameter entry, turn MODE switch ON for 2 seconds to go back to "Review Mode".
(6) Using MODE and RESET switches, show the title (= bdAtA, AL, AnA, or PuLSE).
(7) Turn MODE switch ON for 5 seconds to go back to "Measure Mode".
NOTE: NOTE: For complete detail of MODE and RESET switch operations in steps (1), (2), (6) and (7), see
"Table 8.4 Menu Trees and Switch Operation" on page 24.
8.2.2 Procedure to Enter a Parameter
The procedure to enter a parameter (switch operations sequence in "Parameter Setup Mode") comes in four
ways (numerical value setup, measurement unit setup, decimal point setup and analog 4/20mA trim) that follows:
NOTE: Meanings and functions of individual parameters appear in the Parameter List on pages 16 and 17.
Kind 1 Numerical value setup parameters (F, H, Pu, At, A, A1d, A1H, A2d, A2H, AF, AdAn, and Pon)
The blinking digit in the parameter setup mode is the place of interest.
MODE ... Each time the switch is turned on, the place of interest moves one place to the left.
RESET ... Each time the switch is turned on, the figure in the blinking place increases by one.
Or exponential sign toggles ("E", "-", etc.).
→
Following the parameter setup, hold the MODE switch turned ON for 2 seconds (the new setting is
established and the screen returns to the review mode).
A diagram to show parameter setup flow
Measure Mode
(normal mode)
Review Mode
Parameter Setup
Mode
(4)
(1)
(7)
(6)
(2)
(3)
(5)
Example: Parameter "F" (meter factor)
Digit of interest (blinking)
• Turning MODE "ON" moves the blinking digit one place to the left ("E").
• Turning RESET "ON" increases the figure by one ("2"
→
"3").
Exponential sign (E: 10
+n
, −: 10
−n
)
(The screen above reads F=1.2345×10
+2
L)