12
Cleaning the inside
1) Turn the liner inside out.
2) Hand wash the liner in warm water (
30 °C
) using neutral detergent (e.g. Ottobock 453H10=1
Derma Clean).
3) Thoroughly rinse the liner with clear water until all detergent residues have been removed.
4) Turn the liner again so the outer side/textile material is on the outside.
5) Put a hand towel into the liner and leave the liner to dry lying flat on a drying rack. Alternat
ively, use a drying stand (TT: 5Y100=1 TF: 5Y100=2).
NOTICE! Liner with textile cover: To avoid damaging the liner, dry it with the textile
material on the outside.
Cleaning the outside
►
Clean the outside of the liner approximately every 1 or 2 weeks. The procedure is the same as
for the inside of the liner.
Cleaning the magnetic cap
►
Clean the cap with a dry or moist cloth.
7 Shipment
There must not be any significant magnetic effect on the outside of the packaging when the
product is shipped. The provisions of the shipper must be observed.
1) Keep the original packaging in case the product needs to be shipped.
2)
Tip:
Tilt the packaging at a 45° angle and place a paper clip on it. If the paper clip slides off,
no magnetic effect is present.
8 Technical data
Liners
Size M
Reference number
6Y200=M-1
6Y200=M-2
Length (not shortened, outside dimension) [mm]
450
450
Weight [g]
650
680
Circumference (inside, at the rim of the liner) [mm]
440
470
Circumference (inside, 50 mm from the bottom of the liner) [mm]
330
360
Coupling force to 5A60=* Varos [N]
150 (typical)
Liner without prosthetic socket
Measuring point
Magnetic flux density [mT]
Inside, close to the magnetic cap
<3
Inside, 10 cm from the magnetic cap
<0.05
Outside, close to the magnetic cap
<400
Outside, 10 cm from the magnetic cap
<0.2
Liner with prosthetic socket attached
Measuring point
Magnetic flux density [mT]
Inside, close to the magnetic cap
<15
Inside, 10 cm from the magnetic cap
<0.05
Exterior of prosthetic socket, close to the distal con
nection
<10
Exterior of prosthetic socket, 10 cm from the distal
connection
<0.05