8

4.6 Information on Use

CAUTION

Improper use

Injury due to malfunction and resulting unexpected prosthesis actions.

Instruct the patient in the proper use of the product.

CAUTION

Mechanical stress on the product

Injury due to faulty control or malfunction of the product.

Do not subject the product to mechanical vibrations or impacts.

Check the product for visible damage before each use.

CAUTION

Risk of pinching between the fingertips

Injury due to pinching of body parts.

Ensure that no body parts are between the fingertips when using the product.

CAUTION

Operation outside of the permissible temperature range

Injury due to faulty control or malfunction of the product.

Avoid operation in areas outside the permissible temperature range (see page 13).

CAUTION

Penetration of dirt and moisture in the Axon-Bus components

Injury due to faulty control or malfunction of the Axon-Bus prosthetic system.

Ensure that neither solid particles nor liquids can penetrate into the Axon-Bus prosthetic sys­
tem or Axon-Bus components (e.g. the Axon-Bus gripping component).

CAUTION

Storage of the product in the closed state

Injury  due  to  faulty  control  or  malfunction  of  the  product  as  a  result  of  damage  to  the  sensors
and mechanism.

Store the product only in the neutral position or in the open state.

NOTICE

Improper product care

Damage to the product due to the use of incorrect cleaning agents.

Only clean the product with a damp cloth and mild soap (e.g. 453H10=1-N Ottobock Der­
maClean).

Use only the following products for cleaning/disinfecting the inner socket:

Cleaning:

 453H10=1-N Ottobock DermaClean

Disinfection:

 Colourless, conventional medical disinfectants

Содержание Michelangelo Hand Transcarpal 8E550

Страница 1: ...Instructions for use qualified personnel 3 Michelangelo Hand Transcarpal 8E550...

Страница 2: ...2...

Страница 3: ...GmbH the patient is the operat or of the product according to the IEC 60601 1 2005 A1 2012 standard 2 Product description 2 1 Function The Michelangelo Hand is a myoelectrically controlled gripping c...

Страница 4: ...on setting can be adjusted within a range of 15 see page 11 see fig 8 Opposition Power Grip The opening width makes it possible to hold cylindrical objects with large diameters and with an open thumb...

Страница 5: ...in the case of dysmelia of the forearm The patient must fulfil the physical and mental requirements for perceiving visual acoustic sig nals and or mechanical vibrations 3 4 Contraindications All cond...

Страница 6: ...not open the power supply unit adapter plug or battery charger Do not expose the power supply unit adapter plug or battery charger to extreme loading conditions Immediately replace damaged power supp...

Страница 7: ...to unexpected product behaviour Participation in an Ottobock product training course is mandatory prior to using the product During this product training course you will receive a password giving you...

Страница 8: ...mperature range see page 13 CAUTION Penetration of dirt and moisture in the Axon Bus components Injury due to faulty control or malfunction of the Axon Bus prosthetic system Ensure that neither solid...

Страница 9: ...rging plug on credit cards diskettes audio or video cassettes 5 Scope of Delivery and Accessories 5 1 Scope of delivery 1 pc 8E550 L R M Michelangelo Hand Transcarpal 1 pc 10S550 M lamination ring Inc...

Страница 10: ...nd and gives the Transcarpal prosthesis a natural appearance The recommended alignment is to be used as a guideline for the prosthetic alignment Before the lamination ring is laminated the position of...

Страница 11: ...o short use longer screws Please see the section Technical data for approved screws see page 13 6 1 2 Connecting the Michelangelo Hand with the socket 8 9 10 Once the socket has been completed the rot...

Страница 12: ...riod This is the only way to maintain full warranty cover Additional services such as repairs may be provided in the course of maintenance These addi tional services may be provided free of charge or...

Страница 13: ...on Reference number 8E550 L M or 8E550 R M Opening width 120 mm 4 72 inch Weight of Michelangelo Hand without prosthet ic glove approx 460 g 16 oz Length 12 mm in scope of delivery Length 14 mm in sco...

Страница 14: ...authority pertaining to return and collec tion Declaration of conformity according to the applicable European directives Serial number YYYY WW NNN YYYY year of manufacture WW week of manufacture NNN s...

Страница 15: ...15...

Страница 16: ...th Korea KR 101 131 692 KR 101 169 834 KR 101 178 679 KR 101 178 646 USA US 7 867 287 US 8 016 969 US 8 257 446 US 8 188 835 US 8 579 991 Patents pending in Australia Brazil Canada China EPA Germany I...

Отзывы: